19
NORMALIZATION SETUP
SELECT
SELECT
SELECT
SELECT CHAMBER 1:
ENTER
SELECT
SELECT CHAMBER 2:
SELECT CHAMBER 3:
SELECT CHAMBER 4:
3. Select the appropriate chamber.
CH1 NORM. PARAMETERS 1
SELECT
SELECT
SELECT
Norm. Time: 60 Sec.
Wire diameter: 40 um
Release CH1: 1000 um
ENTER
SELECT
NEXT
SELECT
SELECT
SELECT
Norm. Pressure: 13.3 KPa
Norm. Factor: 0.9
Eyepicece cal.: 0.40 mm/div.
ENTER
BACK
CH1 NORM. PARAMETERS 2
4. Enter the values for the Norm. Time, Wire/Pin diameter.
DMT have the following wire and Pin products:
Steel Wire :
40µm in diameter
Tungsten wires: 25µm, 15µm and 10µm in diameter
Pins:
200µm, 250µm, 300µm and 400µm
in
diameter
5. Press NEXT to go into the PARAMETERS 2 menu
Myograph chamber tubing.