![DMT 630MA User Manual Download Page 15](http://html.mh-extra.com/html/dmt/630ma/630ma_user-manual_2514268015.webp)
15
3.2 MOUNTING PROTOCOL FOR LARGER ARTERIES
For mounting of larger arteries with an internal
diameter larger than 500µm and up to 11mm
replace the mounting Jaws with Pins (see chapter
2.1). With the 630MA system a box with 4 sets of
200µm pins is delivered. For really large arteries
pins with 250µm, 300µm or 400µm pins can be
purchased from DMT.
3.2.1 MOUNTING STEP ONE
• First make sure that the mounting pins is placed correctly across each other in the chambers as
shown below.
Pins from top view
Pins from side view