2.1.2 COARSE ADJUSTING THE JAWS FOR SMALL VESSELS (FIGURE 2.1)
1. Loosen screw A to move the micrometer-side
jaw toward or away from the micrometer.
2. Loosen screw B to move transducer-side jaw
toward or away from the transducer.
3. Loosen screw C to vertically alight the
transducer-side jaw. Screw C is the screw on
the transducer-side support that is furthest
away from the transducer.
NOTE: Number the supports with the number of the chamber they were removed from
using some kind of permanent marker. Store the supports in the provided plastic case.
Numbering the supports will save time when the supports are changed again, limiting
the amount of adjustments needed after each change.