NOTE:
If none of the keys is pressed in the basic mode and in the setting mode, the
device switches off the blue display lighting after approx. 20 seconds. Approx. 1 minute
after the last key has been pressed, a short acoustic signal is generated; after approx. 2
minutes, two short acoustic signals are generated and, after approx. 3 minutes, three
short acoustic signals are generated. Afterwards, the device switches off automatically. If
desired or in an emergency situation, however, you can always switch off the device in
the operating, pause or setting mode at any time by pressing the key
( )
for several
seconds.
Switching off the device:
If the programme time has run out in the operating mode, the device stops the pulse
output after generating an acoustic signal and switches into the basic mode. Press and
hold the key
( )
for several seconds in order to switch off the device completely.
19
1
2
Fig. 17
Fig. 18
4
4
6
6
4
6
6
For the
special programmes for 6 different parts of the body
, press the
corresponding one-touch key (e. g.
H
for hand) in order to call the respective adjustable
user programme. On the left at the bottom of the display, the programme
P01
12.1 Adjustable user programmes
The device is equipped with 5 adjustable user programmes in the
TENS mode
and 9 adjustable user programmes in the
EMS mode
.
Each of the 6 special programmes for 6 different parts of the body
has one adjustable user programme. All these programmes start
with
U
(
U1, U2
etc.)
12.0 Application instructions and mode settings
NOTE:
For safety reasons, the pulse output is reset automatically to
0
if a pulse intensity
of
10 mA
or more has been set and if there is an open circuit. This is for instance the
case if the cables of the electrodes are not connected or damaged. Then, the device
returns to the basic mode.
USAGE
12.2 Calling the adjustable programmes
Switch on the device by press the key ( ) and press the key
(P/II)
in the
TENS mode
or
EMS mode
repeatedly in order
to call the adjustable user programmes until the first of the adjustable user programmes
with the designation
U1
appears on the left at the bottom of the display. By pressing the
key
(P/II)
again, all existing programmes beginning with
U
are called one after the
other. Example: If, for example, a user programme (e. g.
U1
) was set in the TENS
programme, information according to
Fig. P
is shown on the display.
11.9 Pause mode and switching off the device
Pause mode:
If you wish to pause during the application, you can
press the key
(P/II)
in the operating mode in order to access
the pause mode. The pulse output is interrupted and the yellow
indicator lights for channel 1
(CH1)
and/or for channel 2
(CH2)
go out. On the display, the
PAUSE
symbol above and the
symbol
( )
next to the minute information (see
Fig. 17
) will flash.
In order to continue the application, press the key
(P/II)
again
and the device will return to the operating mode and resume the
application.