3.5.4  Temperature correction 

 

 

Enter “System para”-> “Temperature correction” interface   

 

When  you  need  to  view  the  actual  body  surface  temperature  of  the  target  (such  as  the 
measurement test of the certification company), you need to uncheck "Enable" in Temperature 
compensation  of  armpit  on  body  surface  configuration  interface,  and  uncheck  " Automatic 
correction" and "manual correction" in the ambient temperature compensation configuration 
interface. 

 

When you need to view the mapped armpit temperature of the target (such as the actual 
temperature screening application), you need to check "Enable" in Temperature 
compensation of armpit on body surface configuration interface , and check "Automatic 
correction" in the ambient temperature compensation configuration interface, and fill in the 
average indoor ambient temperature (usually 25 

) in the “Environment” 

 

Click "Apply" to complete the setting. 

Attention

 

1. 

The ambient temperature compensation function is used to compensate the temperature 
of the body surface (such as forehead) according to the ambient temperature, so as to 
avoid  inaccurate  temperature  measurement  due  to  too  large  difference  between  the 
actual ambient temperature and the body temperature 

2. 

The  correction  function  of  body  surface’s  armpit  temperature  is  used  to  correct  the 
compensated  body  surface  (such  as  forehead)  temperature  and  convert  it  to  armpit 
temperature of human body, which is more convenient for practical application   

3. 

After the armpit temperature correction of body surface is turned on, the system will 
map  the  body  surface  temperature  between  32 

  ~  35 

  to  the  body  temperature 

between 36.2 

  ~ 36.4 

, so it is normal phenomenon for most people with normal 

body temperature to measure the temperature between 36.2 

  ~ 36.4 

 

Summary of Contents for ThermoVu DTM-700

Page 1: ...ThermoVu DTM 700 Site Survey and Commissioning Guide Digital Ally Inc All rights reserved Document version V0 4 Release date 2020 04 24...

Page 2: ...n direct laser beam or reflection under any circumstances whether the system is turned on or off otherwise it will cause permanent damage to uncooled infrared thermal system This product is a precisio...

Page 3: ...3 1 Computer host requirements 6 3 2 Computer network configuration 6 3 3 Temperature measurement software installation and device connection 7 3 4 Software version of equipment 10 3 5 Client software...

Page 4: ...easurement and screening system belongs to precision temperature measurement and imaging equipment it has the following requirements for engineering exploration application 2 1 Scenario requirements T...

Page 5: ...ible camera lens needs to be adjusted to the maximum to ensure that it is consistent with the infrared thermal camera s field of view Refer to the Appendix for equipment components and assembly instru...

Page 6: ...nstrument If you want to simulate the test of fever personnel you can choose more than 2 normal temperature personnel to test together and one of them can hold a 38 40 warm mineral water bottle and pu...

Page 7: ...ation and device connection 3 3 1 Software installation 1 First Click Install package Inspection System version V X X X X Please run as administrator 2 Second Click Browse button to select software in...

Page 8: ...3 Click Next and Install button to install 4 After installation wait for the system to configure itself then press any key to continue...

Page 9: ...you can see Inspection System software on the desktop Run the software as administrator 3 3 2 Device connection After opening the software with the administrator s permission wait for the device to c...

Page 10: ...needs to be upgraded to IR V3 05 Mar 20 or later Attention The hardware of the infrared camera of the subsequent equipment is changed and the version is shown as V1 01 03_200303 This version does not...

Page 11: ...ference person needs to be standing at a distance of 5m from the camera for calibration standing on the left and right side of the screen in turn First move the yellow calibration points on the infrar...

Page 12: ...and include the whole round target of the blackbody in the blackbody box Then right click the blackbody box to turn on the blackbody and the blackbody temperature should be 37 consistent with the act...

Page 13: ...ll in the average indoor ambient temperature usually 25 in the Environment Click Apply to complete the setting Attention 1 The ambient temperature compensation function is used to compensate the tempe...

Page 14: ...on 3 6 1 Logging in to visible camera Not necessary by default If it is necessary for you to log in to visible camera to check configuration Open IE browser log in IP of 192 168 1 13 then log in user...

Page 15: ...ot alarm 1 Check whether the alarm threshold is set correctly Generally the alarm threshold is set to 37 3 Check whether the alarm threshold is set too large 2 No temperature measurement if no face is...

Page 16: ...e outdoor environment temperature rises sharply and the body surface temperature rises too much with the seasons change it may interrupt the manual compensation of the temperature in the client Genera...

Page 17: ...lid floating point value when running the client At present the client software only adapts to the regional format corresponding to English By default the regional format of the computer is automatica...

Page 18: ...he probability of recurrence of this problem is very low At present there are only avoidance methods but no specific solution which belongs to product defects 4 Appendix 4 1 Equipment components The e...

Page 19: ...pod and tripod accessories For example Note the appearance of tripod actually delivered may be different which does not affect the actual use and the actual arrival shall prevail 2 Take off the fixed...

Page 20: ...is as shown in the figure 5 The distance between camera and blackbody is about 2 5M 4 2 2 Line connection 1 Connect the two ends of the output cable to the camera and DC12V power adapter 2 Plug the tw...

Page 21: ...ware and run it as an administrator 4 3 2 Visible camera software upgrade 1 Open IE browser log in IP of 192 168 1 13 then log in user name of admin when the password is 123456 If the device has been...

Page 22: ...4 3 3 Infrared thermal camera upgrated 1 2 3 4 5 6...

Reviews: