Digicare Biomedical Technology
Operators Manual Multi-parameter Patient Monitor " LifeWindow Lite LW8 "
49
“NIBP LOW MAP” – Indicates that MAP value is lower than the selected MAP LOWER
LIMIT.
“NIBP HIGH MAP” – Indicates that MAP value is higher than the selected MAP
UPPER LIMIT.
Air Leak – Indicates the pressure in the cuff is not reaching the necessary value.
Check if the cuff is properly connected in the hose and in the NIBP
receptacle in the Patient Panel. Check for air leaks in the cuff or hose.
Excessive Motion – Indicates the motion artifact is affecting the NIBP determination.
Time Too Long – Indicates the NIBP cycle is lasting too long for the NIBP determination
and was aborted.
SECTION 8 - TEMPERATURE MONITORING
• The LifeWindow™ Lite LW8 has two temperature channels and one Indicator to T1 – T2
temperature difference. The temperature probes are YSI 400 series compatible.
A. PATIENT CONNECTIONS
• To ensure conformance with all safety and performance specifications, use only the
recommended accessories. These are available from DIGICARE.
Temperature rectal/esophageal probe – disposable
TP001
Temperature skin probe – disposable
TP002
Temperature patient cable for disposable probe
TP003
Temperature rectal/esophageal probe – reusable
TP004
Temperature skin probe – reusable
TP005
Summary of Contents for LifeWindow Lite LW8