B. TEMPERATURE MONITORING
• Connect the temperature patient cable to receptacle (16 ) Temperature 1 or Temperature 2.
• In the Temperature menu select Temperature1, Temperature2 or T1-T2.
• Select Temperature alarm High and Low limits.
• Select Temperature measurement unit: °F or °C.
• The temperature value is displayed when the probe temperature is in the 10 to 50°C (50 to
122 °F), range otherwise the temperature display is blanked.
1 - TEMPERATURE ALARM MESSAGES
TEMP# SENSOR – Indicates the Temperature sensor is not connected to the unit or a sensor
failure.
TEMP# LOW – Indicates the Temperature value is lower than the selected Temperature
Lower Limit.
TEMP# HIGH – Indicates the Temperature value is higher than the selected Temperature
Upper Limit.
SECTION 9 – INVASIVE PRESSURE MONITORING
• The LifeWindow™ Lite LW8 can have 1 or 2 Invasive Pressure Channels installed. The
following information apply for both channels.
• The Invasive Pressure Channels 1 and 2 are turned off in the factory default settings, to avoid
alarms when not in use.
• Turn-on Invasive Pressure Channel 1 or 2 touching in the IBPs menu, then touching Invasive
Pressure 1 or Invasive Pressure 2 and touching IP1 On or IP2 On.
• The message “NO TRANSDUCER” is displayed until a Transducer is connected to the
monitor.
A. PRESSURE TRANSDUCER
To assure the performance specifications in this manual., use only the pressure transducer and
accessories recommended by the manufacturer of this equipment.
IBP cable for reusable transducer
IBP001
IBP reusable transducer
Sensitivity: 50
µ
V / V / mmHg
IBP002
IBP disposable dome for reusable transducer
IBP003