Digicare Biomedical Technology
Operators Manual Multi-parameter Patient Monitor " LifeWindow Lite LW8 "
40
Sidestream Sampling line –
Regular
for MicroFlow™ Capnography
(high humidity)
Part Number: CAP120
Weight: 4.5 grams
Deadspace – adds approximately 7 cc
For use with ET Tube sizes > 4.0 mm
Sidestream Sampling line –
Regular
for MicroFlow™ Capnography
(low humidity)
Part Number: CAP121
Weight: 4.5 grams
Deadspace – adds approximately 7 cc
For use with ET Tube sizes > 4.0 mm
Sidestream Sampling line –
Low Dead Space
for MicroFlow™ Capnography
(high humidity)
Part Number:
CAP122
Weight: 5.8 grams
Deadspace – adds approximately 1 cc
For use with ET Tube sizes < = 4.0 mm
Sidestream Sampling line –
Low Dead Space
for MicroFlow™ Capnography
(low humidity)
Part Number:
CAP123
Weight: 5.8 grams
Deadspace – adds approximately 1 cc
For use with ET Tube sizes < = 4.0 mm
3 – CONNECTION WITH LIFEWINDOW LITE
•
The LifeWindow™ Lite LW8 CO2 function is a Plug and Play technology built in the hardware
and software of the unit.
•
The LifeWindow™ Lite LW8 screen, with internal MicroFlow™ CO2 sidestream module
installed, displays the CO2 indicators and waveform after turned-on. It displays a Pump is Off
message until a Sample Line is connected.
•
The LifeWindow™ Lite LW8 screen, with external MicroFlow™ CO2 sidestream module,
displays the message “CO2 is Off” in the CO2 indicator area and no CO2 trace is present
until the CO2 module is connected. If the CO2 menu or CO2 indicator area is touched in this
situation, nothing happen as the CO2 function is turned off. Insert the MicroFlow™ CO2
sidestream module connector into the receptacle ( 17 ) of the LifeWindow™ Lite LW8. It
displays a Pump is Off message until a Sample Line is connected.
• The LifeWindow™ Lite LW8 MicroFlow™ CO2 sidestream Module has one exhaust port
located in the rear panel of the unit (28) for internal modules or in the external module itself for
attaching scavenging tubing.
•
Insert the sample cell of the sampling kit into the sample cell receptacle of the MicroFlow™
CO2 sidestream Module or in the ( 10 ) receptacle in the LifeWindow™ Lite LW8 as shown. A
“click” will be heard when the sample cell is properly inserted.
Summary of Contents for LifeWindow Lite LW8
Page 1: ......