C. IP SCALE AND LABEL
• Touch in the IBPs menu or in the IBP channel Indicators area to open the desired IP menu
screen.
• The pre-configured default scale for the Invasive Pressure waveform is 0 to +180 mmHg. To
change the scale touch in the IP# Scales drop down menu and touch the desired IP scale.
The Scale options are: -30 to +30 mmHg
0 to +30 mmHg
0 to +60 mmHg
0 to +90 mmHg
0 to +120 mmHg
0 to +180 mmHg
0 to +240 mmHg
0 to +300 mmHg
+30 to +90 mmHg
+60 to +120 mmHg
+90 to +150 mmHg
+120 to +180 mmHg
+150 to +210 mmHg
+180 to +240 mmHg
+210 to +270 mmHg
+240 to +300 mmHg
• The Label displayed in the right side of the Invasive Pressure waveform can be selected to
reflect the site where the Invasive Pressure is being measured.
• The pre-configured default Label is IP1 and IP2 for Invasive Pressure 1 and 2 respectively.
• To change the IBP Label, touch in the IBP# Label drop down menu and select the desired
Label. The Labels are color coded.