![Diesse VES-MATIC EASY Service Manual Download Page 67](http://html.mh-extra.com/html/diesse/ves-matic-easy/ves-matic-easy_service-manual_2495822067.webp)
Service Manual
[TotPagAnnex: 14]
A-5
Rev.1.1 of 15/04/2017
Example of machine status request with address 0x00 without checksum verification.
Example
:
From Host
>0CR+00
The machine with ID ‘00‘ is requested to execute command ‘0x82’ (that is,
command ‘0x02’ without checksum verification)
From V
ES
M
ATIC
E
ASY
>000CR+33
[FREE]
>000CR+33
[ENGAGED]
0x13-P: Request for sending of analyses
As a response to this command, the VesMatic10 instrument will send a 160-byte record containing the
results of 5 tests.
If parameter P is 0x00, the last 5 tests performed will be sent. If parameter P is 0x01, the 5 previous
tests will be sent, and so on. If P is 0x13, the first 5 tests stored in the memory will be sent.
If parameter P exceeds the allowed limits (0x00-0x13), a 1-byte syntax error message (0x01) will be
sent in response.
Example of a request for the 5 tests before the last 5, to the machine with address 0x00 without
checksum verification and with legal parameter P = 0x01.
Example
:
From Host
>000CR+00
The machine with ID ‘00‘ is requested to execute command ‘93’ (that is,
command ‘13’ without checksum verification) with parameter ‘01’
indicating the tests requested.
From VES
>00A00013
[see description below-160 bytes]
+CR+XX
Each 32-byte field containing the results of a test includes the following sub-fields:
E-C-PR-P-T-H-M-S-G-M-A-BC-F1-F2-H1-H2-KI-R1-R2-R3
E (1 byte)
Validates the content of the whole field, if its value is zero. If its value differs from
zero, the remaining bytes in the field have no meaning.
C (1 byte)
Test cycle number.
PR (1 byte)
Contains Boolean flags describing test execution methods:
Bit 0 at 0 means normal test, 1 random test.
Bit 1 at 0 means one hour test, 1 two hour test.
If the test is a 1 h test, sub-fields H2 and KI have no meaning.
Bit 2 not used.
Summary of Contents for VES-MATIC EASY
Page 6: ...Service Manual Rev 1 1 of 15 04 2017 vi 6 76 ...
Page 51: ...Service Manual 51 76 45 Rev 1 1 of 15 04 2017 Electrical diagram Code 20103292 Total pages 3 ...
Page 52: ...Service Manual Rev 1 1 of 15 04 2017 46 52 76 ...
Page 53: ...Service Manual 53 76 47 Rev 1 1 of 15 04 2017 ...
Page 55: ...Service Manual 55 76 49 Rev 1 1 of 15 04 2017 Electrical diagram CODE 20103311 ...
Page 57: ...Service Manual 57 76 51 Rev 1 1 of 15 04 2017 Layout Code 30120000 ...
Page 58: ...Service Manual Rev 1 1 of 15 04 2017 52 58 76 Electrical diagram Code 20103300 ...
Page 59: ...Service Manual 59 76 53 Rev 1 1 of 15 04 2017 7 4 WIRED PHOTOCELL BOARD Layout code 30121610 ...
Page 61: ...Service Manual 61 76 55 Rev 1 1 of 15 04 2017 Electrical diagram CODE 20102462 ...
Page 62: ......