Service Manual
[33/76]
27
Rev.1.1 of 15/04/2017
(The initial operations are carried out incorrectly)
Overall effect
Message on the display
RTC FAILURE (Clock error)
Type of fault
The programme fails to read or update the date and time. There is probably
a communications breakdown between the component housing the
programme and the component that memorises the date and time.
Modules affected
CPU board [4.2.]
Checks recommended
▪
Check the power supplies on the CPU Board.
▪
Reset the CPU Board [Attachment B: Updating of Software]
Corrective Action to be
taken
▪
Replace the CPU Board code 30119871.
Overall effect
Message on the display
UNCALIBRATED (Sensors uncalibrated)
Type of fault
The instrument has lost the adjustment values of the optoelectronic units.
This is probably due to the fact that the programme fails to detect valid
values in the EEPROM. BY pressing
OK
these values will be restored at the
default values unless other problems also exist.
Modules affected
CPU board [4.2.]
Checks recommended
▪
Check the power supplies on the CPU board.
▪
Recalibrate the optoelectronic units and by turning off and turning on the
instrument again, check that the calibration values are saved correctly
[use the Print Settings functions].
▪
Reset the CPU Board [Attachment B: Updating of Software]
Corrective Action to be
taken
▪
Replace the CPU board code 30119871.
Overall effect
Message on the display
EEPROM FAILURE
Type of fault
The instrument has lost the data relating to the analysis memorised in the
archive. This is probably because on turning on the instrument the
programme fails to detect valid values in the archive save on the EEPROM.
By pressing
OK
the archive will be formatted (the Procedure last
approximately 20 seconds).
Modules affected
CPU Board [4.2.]
Checks recommended
▪
Check the power supplies on the CPU board.
▪
Carry out an analysis cycle with valid test tubes. At the end, turn the
instrument off and on again and use the Print Archive function (see
Operating Manual), to check that the analysis results are correctly save in
EErprom.
▪
Reset the CPU Board [Attachment B: Updating of Software]
Summary of Contents for VES-MATIC EASY
Page 6: ...Service Manual Rev 1 1 of 15 04 2017 vi 6 76 ...
Page 51: ...Service Manual 51 76 45 Rev 1 1 of 15 04 2017 Electrical diagram Code 20103292 Total pages 3 ...
Page 52: ...Service Manual Rev 1 1 of 15 04 2017 46 52 76 ...
Page 53: ...Service Manual 53 76 47 Rev 1 1 of 15 04 2017 ...
Page 55: ...Service Manual 55 76 49 Rev 1 1 of 15 04 2017 Electrical diagram CODE 20103311 ...
Page 57: ...Service Manual 57 76 51 Rev 1 1 of 15 04 2017 Layout Code 30120000 ...
Page 58: ...Service Manual Rev 1 1 of 15 04 2017 52 58 76 Electrical diagram Code 20103300 ...
Page 59: ...Service Manual 59 76 53 Rev 1 1 of 15 04 2017 7 4 WIRED PHOTOCELL BOARD Layout code 30121610 ...
Page 61: ...Service Manual 61 76 55 Rev 1 1 of 15 04 2017 Electrical diagram CODE 20102462 ...
Page 62: ......