15
Rev.1.1 of 15/04/2017
-
the maximum stroke set for the reading carriage
-
number of analysis completed by the instrument
-
settings (refer to the Operating Manual)
-
parameters of the QC set (refer to the Operating Manual)
-
the current values of the10 sensors (mA) memorised.
3.1.10. CONFIGURATION OF TEST MODE
[TIME NORMAL/FAST]
This menu allows for reducing the analysis time to the simple mechanical movements of the instrument
and it is therefore possible to rapidly check the settings of the instrument during an analysis cycle.
Procedure:
1.
Select the item ‘TIME NORMAL/FAST’ from the Select test menu,
2. Press
OK
, the mode set will appear on the display:
3. Use the key
▲ to set FAST mode.
4. Press
OK
to save the value entered and return to the previous menu.
These ANALYSIS CYCLES ARE NOT VALID for scientific purposes
Turn off and turn on the instrument again to restore normal functioning.
Method to verify the sensors efficiency using the FAST mode.
1.
To use this rapid procedure it’s necessary to prepare 2 (or 3 for F2 analysis) test-tubes with inside
little sticks (in plastic or in metal) of various heights as shown in figure:
Sample
A
: stick long
56mm
(near the sign level) that simulate a sample
after the mixing.
Sample
B
: stick long
50mm
that simulate a sample after one hour.
Sample
C
: stick long
30mm
that simulate a sample after two hour.
2. Select the ESR F1 ANALYSIS from the
SELECT FUNCTION menu.
3. Introduce the tube
A
and
start a cycle type
F1
.
4. Wait that the instrument executes the two refernce readings, remove the sample
A
and insert the
sample
B
.
5. After the reading at the first hour, the instrument prints the result.
NORMAL
Summary of Contents for VES-MATIC EASY
Page 6: ...Service Manual Rev 1 1 of 15 04 2017 vi 6 76 ...
Page 51: ...Service Manual 51 76 45 Rev 1 1 of 15 04 2017 Electrical diagram Code 20103292 Total pages 3 ...
Page 52: ...Service Manual Rev 1 1 of 15 04 2017 46 52 76 ...
Page 53: ...Service Manual 53 76 47 Rev 1 1 of 15 04 2017 ...
Page 55: ...Service Manual 55 76 49 Rev 1 1 of 15 04 2017 Electrical diagram CODE 20103311 ...
Page 57: ...Service Manual 57 76 51 Rev 1 1 of 15 04 2017 Layout Code 30120000 ...
Page 58: ...Service Manual Rev 1 1 of 15 04 2017 52 58 76 Electrical diagram Code 20103300 ...
Page 59: ...Service Manual 59 76 53 Rev 1 1 of 15 04 2017 7 4 WIRED PHOTOCELL BOARD Layout code 30121610 ...
Page 61: ...Service Manual 61 76 55 Rev 1 1 of 15 04 2017 Electrical diagram CODE 20102462 ...
Page 62: ......