V3.01.51
9
Note: If you get a
“USB Device Not Recognized” pop up. Go to USB
Device Not Recognized in the Troubleshooting section of this
manual.
e) The Add Logger/Repeater window will open. Populate the following fields:
i)
Connect Via:
Tells the Logger where to send its signal. Select
Receiver or an installed Repeater
ii)
Unit type:
Logger
iii)
Unit ID:
Enter any ID you wish (4 digits only (0-9 and A-F)
(1)
Note
: “0000” and “FFFF” are not valid ID’s
(2) Each Repeater/Logger must have a unique ID
iv)
Name:
Any name you wish to give a Logger: Alpha/numeric field 15
characters.
v)
Location:
Location description of where Logger is to be placed:
Alpha/numeric field 20 characters.
vi)
Status:
Active
vii)
Push to Start:
(1)
No:
Logger will automatically start logging once Logger is setup.
(2)
Yes:
Will delay logging till push to start button on Logger is pressed.
Logger Status
Screen will display “no data” until Push-to-Start
button is pressed and the Logger has collected enough data to
send logged data to the Receiver.
Push to start button should
be held down for 2 seconds to activate.
viii)
Sample Rate:
User selectable up to 1 hour starting at 10 seconds in
10 second intervals
ix)
User Calibration Interval:
Defaults to 6 months. Can be set from 1
month to 24 months in 1 month increments.
x)
Audible Alarm On/Off:
Applies to models WT440 and WH445 only.
To reset the audible alarm, hold down the Push-to-Start button for 2
seconds. The alarm reset time depends on the Alarm Delay setting. If
the Alarm Delay is set for 5 minutes, the logger will reset the alarm for
5 minutes, if after 5 minutes the alarm still exist, the logger alarm will
sound.
xi)
Channel Name 1:
25 character name (not required)
xii)
Channel Name 2:
25 character name for WH models only (not
required)
xiii)
Set Email/Text Alert:
Active for 21CFR11 Compliant Secure
WiZARD2.4 Software only
– WR425. Allows up to 6 text/email
messages to issue when alarm condition is met.
(1)
NOTE: Email/Text addresses must first be setup in TOOLS /
EMAIL TEXT ADMIN in order to show in dropdown box.
May be
added to Logger after Logger has been setup. Go to Text/Email
Setup section in this manual for instructions.
xiv)
Alarm Temp/Humidity
: Set min and max alarm conditions for Logger
if desired.
xv)
Alarm Delay:
If desired an alarm delay of up to 999 minutes can be
set.