V3.01.51
22
down the CNTL key while clicking on the desired graph tabs to
highlight then click on Print and select Selected Graphs to save.
(3) To print all open graphs, click on Print and select All Open Graphs
x)
Table:
Will open a table of data points for each open Logger. Will
display date range of data points in the table that are displayed on the
graph.
xi)
Report:
Creates a summary of data shown on the current graph.
(1) Logger Summary: Includes Logger ID, Name, Location, Number of
Data Points Collected, Start & End Date/Time, Total Time
(2) Data Summary: Max, Min, Average, Standard Deviation, Mean
Kinetic Temperature for each active channel.
(3) Alarm Summary: Alarm Min/Max Settings, Time Over Alarm,
Longest Time Over Alarm and Number of Times Over Alarm.
(a) Mean Kinetic Temperature (MKT) formula:
Expressed as:
H/R
-
ln(
e
-
H/RT1
+ e
-
H/RT2
+ …+ e
-
H/RTn
)
( n )
Where
H = activation energy (typically from 60 to 100 kJ/mol for solids and liquids) Use 83.14472
kJ/mol
R = 8.314472 J/mol-K (universal gas constant)
T = temperature in degrees K
n = the number of sample periods over which data is collected
note: ln is the natural log and e
x
is the natural log base.
(b) Standard Deviation (Std. Dev.) formula:
Std Dev = (((R1-Avg)^2 +R2-Avg)^2……(Rn-Avg)^2)/N)^(1/2)
average is already computed and will be labeled Avg.
data points collected is already computed and will be labeled N
the number of each individual reading will be labeled with a suffix n
an individual reading will be labeled R, so a specific nth reading will be labeled Rn
c)
Table Mode:
i)
Right click on Logger in Equipment Status Table and select Table
ii) Multiple tables can be opened and viewed at the same time.
iii) Easy export to Excel
iv) Menu Options: Home, Zoom, Tile, Export, Graph
(1)
Zoom:
(a) Default: The most recent 2 days of data will automatically
display in the table when opened. This default can be changed
by selecting Format/Defaults and changing the range of time to
display.