13
Items of Measure
Frequency
Guideline for
Measurement
Set Value
Item to Check
Water Supply Value
More than once a
day
Design value
Liquid Supply
Pressure
More than once a
day
Design value
Make sure that it does not exceed the maximum water
supply pressure as stipulated in the specifications.
Liquid Supply
Temperature
More than once a
day
Design value
Make sure that it does not exceed the limit of liquid
supply temperature stipulated in the specifications.
Pressure Difference
in Module
More than once a
day
Design value
An increase of more than 0.05MPa (0.5kg/cm2) from the
initial difference pressure may clog or contaminate the
hollow fiber. Either cleanse or replace the module.
Vacuum
More than once a
day
Design value
When degradation of the setup value has occurred,
please check the following 3 points.
•
Air leakage from the vacuum exhaust piping
•
Liquid leakage from vacuum exhaust piping
•
Trouble with the vacuum pump
Confirmation of
Drain (Trap) of
Vacuum Line
More than once a
week
A large amount of liquid may be found in the vacuum
line, which can be considered a leak. In this case, replace
the module.
Dissolved Oxygen
(D/O) in dealt Liquid
More than once a
week
If target D/O value is not attained, check the following.
࣭
D/O in inlet water
࣭
Inlet water volume
࣭
Inlet water temperature
࣭
Vacuum power
0DLQWHQDQFH
It is necessary to have regular checkups to maintain the performance of the degasification
module. Check the items described in the following table and record the results.