8.2.1.1
Short circuit detection
8.2.1.2
Cut circuit detection
8.2.1.3
Pressure error
Binary code
CAUSE
SUGGESTED ACTION
XXX1
Purge valve in short circuit
Change the purge valve
XX1X
Production valve in short circuit
Change the production valve
X1XX
Pump and inlet valves in short circuit
Change the pump and the inlet valves
1XXX
Flush valve in short circuit
Change the flush valve
Binary code
CAUSE
SUGGESTED ACTION
XXX1
Purge valve disconnected or damaged
Check the purge valve
XX1X
Production valve disconnected or
damaged
Check the production valve
X1XX
Pump and inlet valves disconnected or
damaged
Check the pump and the inlet valves
1XXX
Flush valve disconnected or damaged
Check the flush valve
Binary code
CAUSE
SUGGESTED ACTION
XXX1
“
After membrane
”
pressure sensor
reports too low values
Check the pressure sensor after the
membrane
XX1X
“
After membrane
”
pressure sensor
reports too high values
Check the pressure sensor after the
membrane
X1XX
Inlet pressure sensor reports too low
values
-
Check the inlet water
-
Check the inlet pressure sensor
1XXX
Inlet pressure sensor reports too high
values
Check the inlet pressure sensor