Op
erator’s manual
version: 2021.06
13
3.6.7.4
Calibration curve with one point
As the electrodes are not accessible by the user, they are able to measure only the conductivity of the water
crossing the system.
The system uses by default its own calibration curve.
It is recommended to calibrate the 1st point of calibration when installing the system, or when replacing filters,
the RO membrane, resin cartridge, électrodes, or the electronics controller (in all these cases, delete the 2
nd
point).
•
Point of calibration: grey vertical line.
•
The calibration curve goes from 0 to the point of calibration.
•
Above the point of calibration, the values are adjusted by a
shift from the values measured by the electrode.