Op
erator’s manual
version: 2021.06
17
3.6.9.2
To define a 1st point of calibration
This is the default configuration, before any saving of calibration point.
SELECT:PRESS SET 2 SEC
CALIBRATION
PROD CONDUCTIVITY
MEMBRANE CONDUCTIVITY
SELECT:PRESS SET 2 SEC
CONDUCTIVITY CAL
SET MEMBRANE POINT 1
3.6.9.3
To delete the 1st point of calibration
It is possible to come back to the default conductivity values by deleting the 1
st
calibration point.
SELECT:PRESS SET 2 SEC
CALIBRATION
PROD CONDUCTIVITY
MEMBRANE CONDUCTIVITY
SELECT:PRESS SET 2 SEC
DELETE MEMB. POINT 1
CONDUCTIVITY CAL
SET MEMBRANE POINT 1
SELECT:PRESS SET 2 SEC
CONFIRM DELETION OF
MEMBRANE
POINT 1
Delete the point of
calibration by pressing
with a tool 2 seconds on
the ‘SET’ button