Cleaning the device
Personnel: Operating personnel
WARNING!
Ingress of cleaning agents in the device
Electric shock
• Use a moist cloth for cleaning.
Also note the following:
Only clean the control panel with water and detergent. Do not use acetone or solvents. The
consequence would be permanent damage of the plastic surfaces.
Cleaning air-cooled condenser
Personnel: Operating personnel
1.
Switch off the device.
2.
Remove the front cover by holding underneath with both hands and pulling the grating to
the front. Remove the front cover slowly and carefully to prevent damage.
3.
Brush off or vacuum the condenser.
4.
Remount the front cover carefully.
Checking the heat transfer liquid
Soiled or degenerated heat transfer liquid must be replaced.
Replace fuse
1.
Switch off the device.
2.
Remove power cord from the main power socket at the back of the device.
3.
Remove fuse holder from the main power socket at the back of the device.
4.
Replace fuse T10AH (10 ampere time delay fuse with high breaking capacity).
5.
Remount fuse holder in the main power socket.