Equipment installation
The following pages contain information on installing your particular Bioruptor Pico model.
This equipment must only be installed by personnel after reading this section. Consider all
hazards even though no particular hazards have been identified during installation. Before
starting installation work, turn the main switch off and secure the unit against being re-
energized. No special tools are required. One square meter is needed to set-up the Bioruptor
Pico.
Devices are designed to be safe under the following conditions:
Installation overview
• Indoor use
• Altitude up to 2,000 meters
• Operating external temperature 15°C to 25°C
(do not install the Bioruptor in a cold room)
• Maximum relative humidity 80%
• Transient overvoltage typically present on the
MAINS supply
• Degree of protection: IP31
• Power plug must be grounded
• POLLUTION DEGREE 2 (Normally only
non-conductive pollution occurs. However,
occasionally a temporary conductivity caused
by condensation is expected)
• Never install this equipment in a place where
environmental conditions and warnings
mentioned above are infringed
Fig.
Schematic installation overview of the Bioruptor
Pico System in combination with the Bioruptor Cooler.