CHARGING THE BATTERY

 - If not fully charged, the battery will automatically charge when the power unit is 

plugged into a power source via the power adapter. The battery will charge whilst the vacuum power unit is in standby 
or a run mode.

REPLACING THE BATTERY

 - The battery is installed inside the unit and is not accessible by users. Battery 

replacement should only be carried out by qualified service personnel. Contact Direct Healthcare Group or authorised 
dealer for battery replacement service.

6. Care and Maintenance

6.1 Power Unit

Always disconnect the 

VENTURI

 COMPACT power unit from the power adapter and the power adapter from the 

power source before carrying out maintenance, repairs, servicing or cleaning. Check all electrical connections 
and power cord for signs of excessive wear. The power unit / power adapter can be wiped down with detergent or 
disinfectant solution or wipe*. Do not use solvents. Unsuitable for sterilisation. Dispose of the power unit / power 
adapter in accordance with the local regulations including WEEE requirements. The power unit / power adapter 
should be cleaned between patient use as a minimum.

*  In line with the MHRA Medical Device Alert (MDA/2013/019), Direct Healthcare Group advises customers to 
use pH neutral, high level disinfectant cleaning products to sanitise reusable medical devices to prevent damage 
to materials and the degradation of plastic surfaces after prolonged use. The use of inappropriate cleaning and 
detergent materials on medical equipment could damage surfaces and may compromise the ability to decontaminate 
medical devices adequately or may interfere with device function. Direct Healthcare Group recommends the use of 
TECcare

®

 CONTROL antimicrobial wipes and fluid to clean and decontaminate all products it supplies to health and 

social care facilities. TECcare CONTROL products provide class leading broad spectrum, high level disinfection with an 
exceptional safety profile. Being pH neutral TECcare CONTROL can be universally used on all hard and soft surfaces 
without any detrimental effect. TECcare CONTROL is CE marked for cleaning medical equipment.

6.2 Wound Care Sets and Canisters

Wound Care Sets and canisters are disposable and intended for single use only. After use please dispose of in an 
appropriate manner in accordance with local regulations and hospital best practice.

6.3 Servicing

Once the initial guarantee period expires, Direct Healthcare Group recommend that power units should be serviced 
every two years or as indicated by the ‘hours to service’ display.) The unit contains no user serviceable parts and 
should only be serviced by either Direct Healthcare Group or an authorised dealer. Direct Healthcare Group or the 
authorised dealer will make available on request service manuals, component parts lists and other information 
necessary for a competent electrical engineer to repair or service the system. For service, maintenance and any 
questions regarding this, please contact Direct Healthcare Group or an authorised dealer.
It is the customer’s responsibility to ensure the following prior to collection:
•  the system is cleaned of any obvious contaminants.
•  contamination status is documented.
•  assistance is given to Direct Healthcare Group personnel to bag the equipment if the mattress has been in a known 

or suspected infectious environment.

D H G - H E A LT H C A R E . C O M

9

U S E R   M A N U A L

Summary of Contents for TG600/08F

Page 1: ...DHG HE ALTHCAR E C O M VENTURI Compact User Manual...

Page 2: ...VENTURI COMPACT DHG HEALTHCARE COM...

Page 3: ...NPWT 5 Getting Started 5 Operating the Vacuum Power Unit 5 5 Operation Guidelines 6 User Information 6 Operation Controls 7 User Selectable Functions 8 Setting Changing Power Unit Display Languages 8...

Page 4: ...Warning Refer to instruction manual booklet Medical Devices Directive 93 42 EEC Medical Device Regulation 2017 745 North America ETL listed Class II Equipment Double Insulated Do not dispose of with t...

Page 5: ...for closure by primary intention for wound management odour control exudate management etc Wounds which may benefit from the application of NPWT include pressure ulcers dehisced surgical wounds diabet...

Page 6: ...acuum power unit or power adapter on or near a heat source Never use the power adapter whilst placed on top of or near to material which is flammable or can be damaged by heat The supplied power adapt...

Page 7: ...nal s treating the wound must assess how to best use the system for an individual wound 1 Remove all packaging from the power unit and mains adapter NB The power unit internal battery may be mostly di...

Page 8: ...t and continue NPWT 7 Canisters should be replaced as required or weekly To change canister make sure power unit is in stand by mode if still running press and hold the UNLOCK button followed by the R...

Page 9: ...ower unit will then power off after 1 minute of inactivity If using mains power switch off by disconnecting power cable from power unit or turning off mains power VACUUM LEVEL Vacuum level can be adju...

Page 10: ...ey to select this function c To select the required language press and hold the THERAPY MODE button until the language is displayed Releasing the THERAPY MODE button will then set the display for the...

Page 11: ...s adequately or may interfere with device function Direct Healthcare Group recommends the use of TECcare CONTROL antimicrobial wipes and fluid to clean and decontaminate all products it supplies to he...

Page 12: ...he canister is removed both of which cancels the message and sounder and returns the power unit to stand by mode Change canister as detailed on page 6 NB The Canister Full sounder can be silenced by p...

Page 13: ...put Voltage 12V Nominal Vacuum Application Continuous default therapy or intermittent Pressure Range 10 to 180mmHg 0mmHg 20mmHg Fixed Internal Battery 3 7V 10Wh Lithium Ion Rechargeable Cell IP Rating...

Page 14: ...ion Medical electrical equipment safety and essential performance IEC 60601 1 11 Home healthcare environment 8 4 Manufacturer s Guarantee The VENTURI COMPACT power unit is covered by a 24 month manufa...

Page 15: ...e source of the interference by turning nearby or suspect equipment off and see if the interference effects stop In any such event the user is encouraged to try to correct the interference by one of t...

Page 16: ...UT 0 5 cycle At 0 45 90 135 180 225 270 and 315 Mains supply quality for the mains adapter should be that of a typical commercial and or hospital environment In the event of a mains interruption the...

Page 17: ...in all situations Electromagnetic propagation is affected by absorption and reflection from structures objects and people a Field strengths from fixed transmitters such as base stations for radio cell...

Page 18: ...nual was correct at the time of going to press However Direct Healthcare Group reserves the right to modify the specification of any product without prior notice in line with a policy of continual pro...

Page 19: ...DHG HEALTHCARE COM USER MANUAL Notes 1 7...

Page 20: ...et Swatar BKR 4013 Malta Direct Healthcare Group Ltd Withey Court Western Industrial Estate Caerphilly United Kingdom CF83 1BF T 44 0 800 043 0881 F 44 0 845 459 9832 E info directhealthcaregroup com...

Reviews: