DHG TG600/08F User Manual Download Page 9

power unit can also be placed in the carry cases provided. In all instances the power unit/canister assembly should 
be kept upright during use to ensure correct operation - the power unit will display a warning if the unit is over tilted.

Please take care when transporting the device not to drop the system onto a hard surface, bend the canister 
connector or pull on the dressing tube excessively. If the connector is broken a visual and audible alarm will 

be activated and the canister connector should be replaced. Due to contamination the broken connector should be 
disposed of as clinical waste.

5.2 Operation Controls

N.B. Before any operation buttons will function (except MUTE), the power unit must be unlocked* - press and hold the
UNLOCK button until the power unit beeps and the lock symbol clears from the display screen.
The operation buttons on the face of the power unit (Fig. 2) provide the following functions.

D H G - H E A LT H C A R E . C O M

7

U S E R   M A N U A L

RUN/STOP
Press to invoke stand-by mode prior to running power unit. Press again to run power unit. Press and 
hold whilst power unit is running (and unlocked) to cease operation and return to stand-by mode. In 

battery operation the power unit will then power off after 1 minute of inactivity. If using mains power, switch off by 
disconnecting power cable from power unit or turning off mains power.

VACUUM LEVEL
Vacuum level can be adjusted when in stand-by mode and for up to 1 minute after last button 
operation when power unit is running using the UP and DOWN arrow buttons. Pressure can be 

adjusted in 5mmHg increments between 10mmHg and 180mmHg according to treatment requirements, the most 
widely used being between 60-80mmHg. The power unit will begin operation at the default pressure of 80mmHg. The 
selected vacuum level is shown on the display screen.

MUTE
Press to silence the sounder and to clear the message from the display screen. The MUTE button can also 
be used to force power-off in stand-by mode.

THERAPY MODE
The power unit offers a choice of continuous or intermittent therapy modes. On power-on, the power unit 
will default to continuous therapy mode. To switch between therapy modes, press and hold the THERAPY 

MODE button until power unit beeps to confirm change of mode. When in intermittent therapy mode, the power unit 
will provide vacuum therapy for periods of 5 minutes followed by a 2 minute rest period. The selected therapy mode 
is displayed on the power unit screen and can be altered in either stand-by or (unlocked) run modes.

Fig. 2

RUN/

STOP

VACUUM 

LEVEL

MUTE

THERAPY 

MODE

UNLOCK/ 

LIGHT

Summary of Contents for TG600/08F

Page 1: ...DHG HE ALTHCAR E C O M VENTURI Compact User Manual...

Page 2: ...VENTURI COMPACT DHG HEALTHCARE COM...

Page 3: ...NPWT 5 Getting Started 5 Operating the Vacuum Power Unit 5 5 Operation Guidelines 6 User Information 6 Operation Controls 7 User Selectable Functions 8 Setting Changing Power Unit Display Languages 8...

Page 4: ...Warning Refer to instruction manual booklet Medical Devices Directive 93 42 EEC Medical Device Regulation 2017 745 North America ETL listed Class II Equipment Double Insulated Do not dispose of with t...

Page 5: ...for closure by primary intention for wound management odour control exudate management etc Wounds which may benefit from the application of NPWT include pressure ulcers dehisced surgical wounds diabet...

Page 6: ...acuum power unit or power adapter on or near a heat source Never use the power adapter whilst placed on top of or near to material which is flammable or can be damaged by heat The supplied power adapt...

Page 7: ...nal s treating the wound must assess how to best use the system for an individual wound 1 Remove all packaging from the power unit and mains adapter NB The power unit internal battery may be mostly di...

Page 8: ...t and continue NPWT 7 Canisters should be replaced as required or weekly To change canister make sure power unit is in stand by mode if still running press and hold the UNLOCK button followed by the R...

Page 9: ...ower unit will then power off after 1 minute of inactivity If using mains power switch off by disconnecting power cable from power unit or turning off mains power VACUUM LEVEL Vacuum level can be adju...

Page 10: ...ey to select this function c To select the required language press and hold the THERAPY MODE button until the language is displayed Releasing the THERAPY MODE button will then set the display for the...

Page 11: ...s adequately or may interfere with device function Direct Healthcare Group recommends the use of TECcare CONTROL antimicrobial wipes and fluid to clean and decontaminate all products it supplies to he...

Page 12: ...he canister is removed both of which cancels the message and sounder and returns the power unit to stand by mode Change canister as detailed on page 6 NB The Canister Full sounder can be silenced by p...

Page 13: ...put Voltage 12V Nominal Vacuum Application Continuous default therapy or intermittent Pressure Range 10 to 180mmHg 0mmHg 20mmHg Fixed Internal Battery 3 7V 10Wh Lithium Ion Rechargeable Cell IP Rating...

Page 14: ...ion Medical electrical equipment safety and essential performance IEC 60601 1 11 Home healthcare environment 8 4 Manufacturer s Guarantee The VENTURI COMPACT power unit is covered by a 24 month manufa...

Page 15: ...e source of the interference by turning nearby or suspect equipment off and see if the interference effects stop In any such event the user is encouraged to try to correct the interference by one of t...

Page 16: ...UT 0 5 cycle At 0 45 90 135 180 225 270 and 315 Mains supply quality for the mains adapter should be that of a typical commercial and or hospital environment In the event of a mains interruption the...

Page 17: ...in all situations Electromagnetic propagation is affected by absorption and reflection from structures objects and people a Field strengths from fixed transmitters such as base stations for radio cell...

Page 18: ...nual was correct at the time of going to press However Direct Healthcare Group reserves the right to modify the specification of any product without prior notice in line with a policy of continual pro...

Page 19: ...DHG HEALTHCARE COM USER MANUAL Notes 1 7...