DHG TG600/08F User Manual Download Page 10

*UNLOCK / LIGHT
The power unit will automatically lock 1 minute after last button operation when running to prevent 
inadvertent operation of button functions (except MUTE), as indicated by    on the display screen. Press 

and hold the UNLOCK button until power unit beeps if further button operation is needed (i.e. change of therapy 
mode or vacuum level, or returning power unit to stand-by mode). The power unit will lock again 1 minute after last 
button operation. Pressing this button will also illuminate the display screen for 10 seconds if power unit is in battery 
operation (screen will always be illuminated in mains operation).

5.3 User Selectable Functions

The following functions are selectable in a user menu:-
•  Language selection
•  Tilt indicator, switch on or off
•  Tube constriction indicator, switch on or off
•  Battery inert. Note that it is essential that the battery is put to this condition before international transportation.
The menu is accessed by a three second press of the UNLOCK button. The arrow buttons scroll to the functions 
shown above, and the THERAPY MODE button allows the function to be toggled on or off, or selected. Pressing 
the UNLOCK button returns the device to stand-by mode.

5.4 Setting / Changing Power Unit Display Language

The power unit display default language is English, but can easily be changed to the language of your choice, as 
follows:-
a)  Access the user menu by pressing the UNLOCK 

button for three seconds.

b)  Scroll through the user functions using the arrow 

buttons until ‘Set Language’ is displayed (Fig. 
3). Press the THERAPY MODE key to select this 
function.

c)  To select the required language, press and hold 

the  THERAPY MODE button until the language 
is displayed. Releasing the THERAPY MODE 
button will then set the display for the language. 
Pressing the UNLOCK button returns the device to 
stand-by mode.

5.5 Battery Information

•  A fully charged battery should operate the power unit continuously for at least 24 hours.
•  Charge status is shown on the display of the power unit when it is in stand-by and run mode.
•  As the batteries are automatically charged as required when the system is operating on mains power, the battery module 

should not require removing or changing in normal use.

•  Use only the mains adapter supplied with the system.
•  When power unit operation times when running from the internal battery are noticeably shorter than normal, it is time to 

replace the battery pack. Contact Direct Healthcare Group or authorised dealer for battery replacement service.

•  Never use any battery pack that is damaged or worn out. Use the battery pack only for its intended purpose.
•  The battery pack is not serviceable and should be replaced if faulty (indicated by applicable battery faults displayed in 

place of battery charge status icon). Contact Direct Healthcare Group or authorised dealer for battery replacement service.

•  The internal Lithium Ion Battery may be received or become inert if left for extended periods in storage or a fully discharged 

condition. Connection of a power adapter will wake / charge the battery and immediately allow the full use of the vacuum 
power unit. It is possible to force the vacuum power unit into making the battery inert should this be required by some airline 
and other carriers. See ‘User Selectable Functions’ Menu for details (Page 8).

D H G - H E A LT H C A R E . C O M

8

V E N T U R I

®

  C O M PA C T

Fig. 3

Summary of Contents for TG600/08F

Page 1: ...DHG HE ALTHCAR E C O M VENTURI Compact User Manual...

Page 2: ...VENTURI COMPACT DHG HEALTHCARE COM...

Page 3: ...NPWT 5 Getting Started 5 Operating the Vacuum Power Unit 5 5 Operation Guidelines 6 User Information 6 Operation Controls 7 User Selectable Functions 8 Setting Changing Power Unit Display Languages 8...

Page 4: ...Warning Refer to instruction manual booklet Medical Devices Directive 93 42 EEC Medical Device Regulation 2017 745 North America ETL listed Class II Equipment Double Insulated Do not dispose of with t...

Page 5: ...for closure by primary intention for wound management odour control exudate management etc Wounds which may benefit from the application of NPWT include pressure ulcers dehisced surgical wounds diabet...

Page 6: ...acuum power unit or power adapter on or near a heat source Never use the power adapter whilst placed on top of or near to material which is flammable or can be damaged by heat The supplied power adapt...

Page 7: ...nal s treating the wound must assess how to best use the system for an individual wound 1 Remove all packaging from the power unit and mains adapter NB The power unit internal battery may be mostly di...

Page 8: ...t and continue NPWT 7 Canisters should be replaced as required or weekly To change canister make sure power unit is in stand by mode if still running press and hold the UNLOCK button followed by the R...

Page 9: ...ower unit will then power off after 1 minute of inactivity If using mains power switch off by disconnecting power cable from power unit or turning off mains power VACUUM LEVEL Vacuum level can be adju...

Page 10: ...ey to select this function c To select the required language press and hold the THERAPY MODE button until the language is displayed Releasing the THERAPY MODE button will then set the display for the...

Page 11: ...s adequately or may interfere with device function Direct Healthcare Group recommends the use of TECcare CONTROL antimicrobial wipes and fluid to clean and decontaminate all products it supplies to he...

Page 12: ...he canister is removed both of which cancels the message and sounder and returns the power unit to stand by mode Change canister as detailed on page 6 NB The Canister Full sounder can be silenced by p...

Page 13: ...put Voltage 12V Nominal Vacuum Application Continuous default therapy or intermittent Pressure Range 10 to 180mmHg 0mmHg 20mmHg Fixed Internal Battery 3 7V 10Wh Lithium Ion Rechargeable Cell IP Rating...

Page 14: ...ion Medical electrical equipment safety and essential performance IEC 60601 1 11 Home healthcare environment 8 4 Manufacturer s Guarantee The VENTURI COMPACT power unit is covered by a 24 month manufa...

Page 15: ...e source of the interference by turning nearby or suspect equipment off and see if the interference effects stop In any such event the user is encouraged to try to correct the interference by one of t...

Page 16: ...UT 0 5 cycle At 0 45 90 135 180 225 270 and 315 Mains supply quality for the mains adapter should be that of a typical commercial and or hospital environment In the event of a mains interruption the...

Page 17: ...in all situations Electromagnetic propagation is affected by absorption and reflection from structures objects and people a Field strengths from fixed transmitters such as base stations for radio cell...

Page 18: ...nual was correct at the time of going to press However Direct Healthcare Group reserves the right to modify the specification of any product without prior notice in line with a policy of continual pro...

Page 19: ...DHG HEALTHCARE COM USER MANUAL Notes 1 7...

Page 20: ...et Swatar BKR 4013 Malta Direct Healthcare Group Ltd Withey Court Western Industrial Estate Caerphilly United Kingdom CF83 1BF T 44 0 800 043 0881 F 44 0 845 459 9832 E info directhealthcaregroup com...

Reviews: