18
12. Use the
↑
or
↓
keys to select the time (in seconds) that the unit will wait before
storing the current measurement and advancing to the next measurement position.
Press
ENTER
to confirm your selection and the display will then indicate:
GOOD MEAS DELAY
1.0 sec
POOR APPL DELAY
2.0 sec
13. Use the
↑
or
↓
keys to select the time (in seconds) that the unit will wait after a
poor applanation before advancing to the next measurement position. Press
ENTER
to confirm your selection.
At this point, the
Pachette 2
will either exit the configuration menu or advance to the
next configurable parameter. More parameters will only be available if other options
have been installed.
Note 1:
To save a changed parameter and immediately exit the configuration menu
without advancing to the next parameter, press the
CLEAR
key.
Note 2:
When changing from Continuous Avg to Mapping mode, any measurements
obtained in Continuous Avg mode will be deleted before the Mapping mode becomes
active. This is necessary because the two operational modes are very different and
measurements from the previous mode would have no relevance if carried over to the
new mode.