11
Table 1
Parameters On The Configuration Menu And Associated Options
Parameter
[ default value ]
Range
Description
Associated
Option(s)
OPERATIONAL MODE
[ Continuous Avg ]
Continuous Avg
Mapping
Selects how the
Pachette 2
should operate.
Continuous Avg = multiple measurements at a single
location.
Mapping = single measurement at different positions.
Continuous
Avg with
Mapping
STD. DEVIATION
[Enabled]
Enabled
Disabled
Enables or Disables the presentation of standard
deviation on the display.
Continuous
Avg
NUMB OF MEAS
[ 25 ]
1 to 50
Selects the number of measurements to be obtained.
Continuous
Avg
AUTO REP DELAY
[ 100 msec ]
100 to 950 msec
Period of time between consecutive measurements
while the probe is applanated to the cornea.
Continuous
Avg
NUMB OF POSN
[ 33 ]
1 to 33
Selects the number of positions to be obtained.
Mapping
AMOUNT OF BIAS
[ 100% ]
1 to 199
Selects the percentage used to calculate biased
measurements.
Mapping
GOOD MEAS DELAY
[ 1.0 sec ]
1.0 to 9.5
seconds
Period of time before the unit will automatically store
the current measurement and advance to the next
measurement position.
Mapping
POOR APPL DELAY
[ 2.0 sec ]
1.0 to 9.5
seconds
Period of time after a poor applanation occurs before
the unit will automatically advance to the next
measurement position.
Mapping
EXT. INTERFACE
[ Disabled ]
Disabled
Prnt Enabled
Xfer Enabled
IrDA Enabled
Mode of operation for the external interface:
Disabled
= no interface is enabled.
Prnt Enabled = RS232C serial connection to
printer.
Xfer Enabled = RS232C serial data transfer to a
computer.
IrDA Enabled = wireless connection for IrDA
device.
Printer
Data Transfer
IrDA
BAUD RATE
[ 9600 bps ]
300
600
1200
2400
4800
9600
Sets the data transfer rate in bits per second (bps) for
the RS232C serial interface.
Printer
Data Transfer
IrDA
PRINT NOTES
[ Enabled ]
Enabled
Disabled
Enables or Disables the Notes box on printout.
Printer