8
V. CONFIGURING THE PACHETTE 2
Configuration Overview
When shipped from the factory, the
Pachette 2
is ready to take corneal measurements.
It is not necessary to setup or configure anything. The operator only needs to connect
the probe to the front panel and then apply power to the unit. However, the
Pachette 2
has been designed to allow the operator to modify certain default parameters to tailor
the instrument to meet one’s needs. Once modified, these parameters are permanently
stored in non-volatile memory and are automatically recalled each time the unit is
powered up. To change a parameter, the operator must access the configuration menu.
The following procedure explains how to access the configuration menu and modify the
default parameters.
Changing Parameters On The Configuration Menu
A complete list of all the possible parameters on the configuration menu is given in
Table 1. The unshaded sections of this table identify parameters that are accessible on
the basic
Pachette 2
. The shaded sections of the table identify parameters that are only
accessible when optional features have been added to the unit. The procedure below
shows how to activate the configuration menu and modify the parameters on the basic
Pachette 2
. For a detailed description of any parameters associated with optional
features, refer to section IX.
1. Press
the
CONFIG
key and the display will briefly indicate:
then the display will briefly show the Software Version and then show:
PACHETTE 2
CONFIG
OPERATIONAL MODE
Continuous Avg
2. Press
ENTER
to continue in Continuous Avg mode or press the
↑
or
↓
key to
change the operational mode from Continuous Avg to Mapping. Refer to section
IX.A to continue in Mapping Mode. In Continuous Avg mode display will show: