4.3.1.3.1.1. IOLITE-GATE: Power in: Pinout
Power in connector: pin-out (2-pin LEMO male)
Pin
Name
Description
1
V +
Supply
2
V -
Ground
For the power supply an unregulated DC voltage
between 12 V and 48 V is required, which is
connected to the 2-pin LEMO 1B male connector
on the IOLITE-GATE front.
PWR IN connector (on the device): EXJ.1B.302.HLD
Mating connector (for the cable): FGG.1B.302.CLAD52Z
4.3.1.3.1.2. IOLITE-GATE: Sync: Pinout
The sync connectors are required when you want to synchronize the data from IOLITE with Dewesoft
USB devices for the same measurement. The signal that is transferred over sync cable makes sure that
the measurement data of IOLITE and Dewesoft USB devices are perfectly synchronized to each other.
The other use of a sync connector is to connect directly to IOLITE, a signal from the clock master.
Hint
There is no distinction between the IN and OUT – it does not matter which connector you use.
When IRIG-synchronisation is used, the IRIG signal is on pins 1, 2.
Sync connector: pin-out (4-pin LEMO female)
Pin
Name
Description
1
CLK
Clock
2
TRIG
Trigger
3
RES
PPS
4
GND
Ground
SYNC connector (on the device): EEG.00.304.CLL
Mating connector (for the cable): FGG.00.304.CLAD27Z
IOLITE® V20-1
35/153
Summary of Contents for IOLITE
Page 1: ...IOLITE TECHNICAL REFERENCE MANUAL IOLITE V21 1 1 ...
Page 14: ...IOLITE TECHNICAL REFERENCE MANUAL Channel setup IOLITE R8 IOLITE V20 1 14 153 ...
Page 41: ...IOLITE TECHNICAL REFERENCE MANUAL IOLITE R8 Back panel IOLITE V20 1 41 153 ...
Page 112: ...IOLITE TECHNICAL REFERENCE MANUAL 5 5 5 2 5 STG Resistance IOLITE V20 1 112 153 ...