IOLITE®
TECHNICAL REFERENCE MANUAL
IOLITEi-8xTH module
5.7.1. TH: Specifications
Inputs
Input type
Isolated universal thermocouple and voltage
Number of channels
8
ADC Type
24-bit delta-sigma
Sampling Rate
Simultaneous 100 S/sec
Voltage Ranges
±1 V
±100 mV
Input Accuracy
±0.02 % of reading ±100 μV
±0.02 % of reading ±10 μV
Typical Noise floor @ 10/100 S/sec
-114 dB / -105 dB
-109 dB / -100 dB
Gain Drift
Typical 4 ppm/K (max. 10 ppm/K)
Offset Drift
Typical 0.05 μV/K (max. 0.2 μV/K)
Gain Linearity
<0.01%
Input Coupling
DC
Input Impedance
100 MΩ
Thermocouple
TC Types: K, J, T, R, S, N, E, C, B
Accuracy
±0.02 % of reading ±0.5 °C ±10 μV
Resolution
< 0.001 °C
Sampling rates
10, 20, 40, 80, 100 S/sec
Typical Noise
0.007 °C RMS@Type K @ 10 S/sec
0.02 °C RMS@Type K @ 100 S/sec
IOLITE® V20-1