2.1 Table design 

When designing the table frames, special emphasis was placed on functional and operational safety. The 
number of possible pinching points has thus been minimised, while remaining ones have been covered or 
protected with spacers to prevent injury, thus ensuring safe and yet simple operation. Nevertheless, 
necessary caution must always be exercised when using the table. 

 

 

The table consists of the following assemblies: - underframe, - scissor section, - upper frame, - upholstery. 
Depending on the respective version, these assemblies can feature further attachment parts. The surfaces of 
the welded design are plastic coated. 
The height adjustment unit is located between the scissor section and the underframe, which guarantees very 
high power transmission even in the lowest adjustment range (min. height). By extending or retracting the 
lifting tube, the scissors are pushed apart or together, thus enabling adjustment of the lying surface.  
The electrical adjustment system does not pose a hazard to the health and safety of the user or the patient 
when used as intended. The lifting motor is activated by a low control voltage. 
 
 

2.2 Height adjustment 
 
Height adjustment (all models with electromotive height adjustment) 

To adjust the height, the enclosed foot switch (optionally also hand switch) is operated according to the 
marking.  
Beforehand, brief activation (double tap) must take place via the foot switch (or the hand switch). Please refer 
to section 7. The table is lifted or lowered. 

 

 

Height adjustment using foot switch rails (optional) 

The electric motor for height adjustment is operated by a switch rail attached to the long side of the table, 
which can be operated with your foot.  
Press the switch rail down = the table is lifted.  
Lift the switch rail up = the table is lowered.  
Here, too, brief activation (double tap) must take place beforehand (see section 7). 
Alternatively, the switch rails can also be led out to the short side of the table. This facilitates  
height adjustment from the short sides of the table. The lifting motor is equipped with a freewheeling clutch as 
standard. This ensures automatic declutching if an obstacle is encountered when lowering. In other words, 
the active tractive force of the motor no longer acts; instead, simply the weight of the upper part of the table is 
applied. In the event of unforeseen entrapment, the risk of injury is significantly reduced.  
 

Note on operation 
The electric motor is to be operated in intermittent duty mode. This means that a maximum duty cycle 
of 25 s must not be exceeded. Before switching the motor back on, an interruption time of at least 400 
s must be observed. If the maximum duty cycle is exceeded, an internal thermal switch (protective 
temperature limiter) in the motor interrupts the power supply to the actuator. After the electric motor 
has cooled down, the thermal switch automatically reconnects the power supply to the actuator. 
Height adjustment (with hydraulic height adjustment) 
 

At tables equipped with hydraulic height adjustment, the height is adjusted by repeatedly depressing 
(pumping) the foot lever on one side of the table. 
To lower the table, the foot lever is simply lifted with your foot.  
If the table is only lifted slightly each time the foot lever is depressed after transport or a longer period of non-
use, air bubbles have formed in the hydraulic system. To remove the air bubbles, pump the table upwards 
under load and perform an additional 20 - 30 strokes of the pump when the table is in the uppermost position. 
This will force air out of the system. 
 
 

2.3 Upholstery adjustment 

 

 

Despite the very robust and strong design of the gas spring, the head part must not be used for 
sitting! 
 
 

Summary of Contents for 1000/H

Page 1: ...pliers guarantees years of trouble free operation provided the device is handled correctly and as intended in accordance with the conditions described in the user manual In the unlikely event you need...

Page 2: ...al number 5 2 Operating manual 6 2 1 Table design 6 2 2 Height adjustment 6 2 3 Upholstery adjustment 6 2 4 Movability model dependent or optional 7 2 5 Additional equipment 7 3 Additional accessories...

Page 3: ...ing the device We recommend that you keep these instructions for use near the device for future reference The device may only be used by authorised instructed and competent persons who are sufficientl...

Page 4: ...level flat and solid surface Before use activate the brakes on the castors and make sure they are working properly 1 6 Commissioning The device is ready for use upon delivery Remove the power cable fr...

Page 5: ...nly take hold of the underframe NOT the upper frame 1 8 Model designation and type labelling The exact model designation depends on the choice of frame colour 00 white powder coated RAL 9010 03 white...

Page 6: ...ss the switch rail down the table is lifted Lift the switch rail up the table is lowered Here too brief activation double tap must take place beforehand see section 7 Alternatively the switch rails ca...

Page 7: ...table feet all four castors are activated simultaneously The following moving options exist Stage 1 The castors of the table are locked and can neither be moved nor rotated Stage 2 centre position The...

Page 8: ...de guard with the necessary caution Never operate the side guard if the hands fingers etc of another person are located between the bars or on the mechanism of the side guard Risk of being crushed pin...

Page 9: ...dle part length mm 480 480 Foot part length mm 920 920 Weight approx depending on equipment kg 90 90 Min max height mm 500 to 940 490 to 940 Adjustment time motor s 22 Head part adjustment range 40 30...

Page 10: ...ion 3 7 A Protection class II protective insulation connecting cable without protective conductor Protection rating IPX4 resistant to water splashes Degree of protection B The motor is maintenance fre...

Page 11: ...nt block the foot levers The power plug must not come into contact with water or cleaning agents The electrical components must not display any external damage through which liquid could enter Do not...

Page 12: ...use Both can be replaced without too much effort Please request the corresponding installation diagram if required Replacement parts can be obtained directly from Dewert 7 Safety devices Tables with...

Page 13: ...ckaging Safety notice Pay attention to sharp edges and pointed objects during disposal The packaging materials produced are mainly o Cardboard Paper o Plastic o Wood when delivered on a pallet Please...

Page 14: ...ith the options of movability not wheel lifting fixing mechanism side guards and push handle the intended purpose of the table is extended and also provides for the ideal positioning of patients for t...

Page 15: ...damaged All switches and supply lines undamaged Function check With electromotive adjustment Move all motors with the foot switch or the hand switch to both limit positions until automatic switch off...