Dewert 1000/H User Manual Download Page 16

OK

FAIL

Description of defects

Height retained?
Hydraulic pump leaking?
Smooth lowering possible?
Check by operating the pedals to lower the table
Strong noise development?
Wipe the piston rod with a cloth

Lying surface adjustment functions:

Metal ratchets - arrester - gas spring

Metal ratchet inspection:  Lifting the lying surface segment:

Do the two metal ratchets engage properly and securely?
Do they engage evenly?
Is this ensured in every adjustment position?

Arrester inspection:  Lifting the lying surface segment:

Is the segment held properly and securely at every height?
Perform the test also with a load
Is the function smooth without clamping?
(= move the lying surface segment without fixing the clamping lever)

Gas spring inspection: Lifting the lying surface segment:

Does the gas spring respond when released?

 

Is the segment held properly and securely at every height?
Is the piston rod of the gas spring free of grease and not leaking?
Clean the piston rod with a cloth

Accessories:

Accessories such as belts, belt pads, belt guides, paper roll holders, 
armrests, etc. undamaged, and is proper and secure fixing/function possible?
All necessary toggle screws present?  

Possible movability:

Castors undamaged, freewheeling given?
Connection to frame undamaged?
Re-tighten all screw connections (with central locking system,
also the grub screws of switch levers)
Safe braking effect?
Check with locked brakes by pulling and pushing the table

Structural inspection:

Checking the scissor fittings:
Use a wrench to check the six fixing screws of the scissors (remove the black caps for this purpose)
and the hexagon socket head screws for a very tight fit

Checking the side guard:

Easy adjustment possible without clamping?
Proper locking during setup/lowering?

Laterally lowerable side guard:

No adjustment possible without pulling the locking knob?
Proper engagement in the end positions?

Electrical testing

Leakage current measurement (protection class II, degree of protection B)
(max. 0.1 mA permissible)

Measured value:

Final assessment

Everything error free?

Device put out of operation until it has been repaired?

Comments

Place / Date / Signature of inspector

Next inspection

 

Summary of Contents for 1000/H

Page 1: ...pliers guarantees years of trouble free operation provided the device is handled correctly and as intended in accordance with the conditions described in the user manual In the unlikely event you need...

Page 2: ...al number 5 2 Operating manual 6 2 1 Table design 6 2 2 Height adjustment 6 2 3 Upholstery adjustment 6 2 4 Movability model dependent or optional 7 2 5 Additional equipment 7 3 Additional accessories...

Page 3: ...ing the device We recommend that you keep these instructions for use near the device for future reference The device may only be used by authorised instructed and competent persons who are sufficientl...

Page 4: ...level flat and solid surface Before use activate the brakes on the castors and make sure they are working properly 1 6 Commissioning The device is ready for use upon delivery Remove the power cable fr...

Page 5: ...nly take hold of the underframe NOT the upper frame 1 8 Model designation and type labelling The exact model designation depends on the choice of frame colour 00 white powder coated RAL 9010 03 white...

Page 6: ...ss the switch rail down the table is lifted Lift the switch rail up the table is lowered Here too brief activation double tap must take place beforehand see section 7 Alternatively the switch rails ca...

Page 7: ...table feet all four castors are activated simultaneously The following moving options exist Stage 1 The castors of the table are locked and can neither be moved nor rotated Stage 2 centre position The...

Page 8: ...de guard with the necessary caution Never operate the side guard if the hands fingers etc of another person are located between the bars or on the mechanism of the side guard Risk of being crushed pin...

Page 9: ...dle part length mm 480 480 Foot part length mm 920 920 Weight approx depending on equipment kg 90 90 Min max height mm 500 to 940 490 to 940 Adjustment time motor s 22 Head part adjustment range 40 30...

Page 10: ...ion 3 7 A Protection class II protective insulation connecting cable without protective conductor Protection rating IPX4 resistant to water splashes Degree of protection B The motor is maintenance fre...

Page 11: ...nt block the foot levers The power plug must not come into contact with water or cleaning agents The electrical components must not display any external damage through which liquid could enter Do not...

Page 12: ...use Both can be replaced without too much effort Please request the corresponding installation diagram if required Replacement parts can be obtained directly from Dewert 7 Safety devices Tables with...

Page 13: ...ckaging Safety notice Pay attention to sharp edges and pointed objects during disposal The packaging materials produced are mainly o Cardboard Paper o Plastic o Wood when delivered on a pallet Please...

Page 14: ...ith the options of movability not wheel lifting fixing mechanism side guards and push handle the intended purpose of the table is extended and also provides for the ideal positioning of patients for t...

Page 15: ...damaged All switches and supply lines undamaged Function check With electromotive adjustment Move all motors with the foot switch or the hand switch to both limit positions until automatic switch off...

Page 16: ...ssories Accessories such as belts belt pads belt guides paper roll holders armrests etc undamaged and is proper and secure fixing function possible All necessary toggle screws present Possible movabil...

Reviews: