Dewert 1000/H User Manual Download Page 15

Checklist for maintenance/technical inspection

to IEC 62353 and DGUV Regulation 3

Device
Model designation
Serial no.
Location
Responsible person
Date, inspector
Actuator designation

Inspections

OK

FAIL

Description of defects

Visual inspection

Overall impression of the table OK?

Labels, CE mark, type plate present?

Manufacturer’s operating manual available and accessible?

Sufficient space available when carrying out all adjustment functions?

Mechanical structure undamaged:

Welds without obvious damage?

Screw connections correct and complete?

Upholstery undamaged? 
Upholstery attachment correct?

All mechanical elements intact and complete?

Electrical system and power cable undamaged?

All switches and supply lines undamaged?

Function check

With electromotive adjustment:

Move all motors with the foot switch or the hand switch to both limit positions until 
automatic switch-off to ensure that:
* the table operates smoothly without any collisions or blockages
* no cable/connection can be overstretched, crushed or otherwise
   damaged
* the motors run without noticeable noise development
* the end position switch-off of the actuators works properly
Additionally:
Power cables and power plugs undamaged?

Correct and safe routing of the power cable and connection?

Checking the safety device:
Double-tap function and freewheeling motor, function given?

With hydraulic height adjustment:

Function given? 
Check by operating the pedals until the table reaches the uppermost position
Continue to operate the pedals approx. 5-10 times (any air is pressed out of the system)

 

Summary of Contents for 1000/H

Page 1: ...pliers guarantees years of trouble free operation provided the device is handled correctly and as intended in accordance with the conditions described in the user manual In the unlikely event you need...

Page 2: ...al number 5 2 Operating manual 6 2 1 Table design 6 2 2 Height adjustment 6 2 3 Upholstery adjustment 6 2 4 Movability model dependent or optional 7 2 5 Additional equipment 7 3 Additional accessories...

Page 3: ...ing the device We recommend that you keep these instructions for use near the device for future reference The device may only be used by authorised instructed and competent persons who are sufficientl...

Page 4: ...level flat and solid surface Before use activate the brakes on the castors and make sure they are working properly 1 6 Commissioning The device is ready for use upon delivery Remove the power cable fr...

Page 5: ...nly take hold of the underframe NOT the upper frame 1 8 Model designation and type labelling The exact model designation depends on the choice of frame colour 00 white powder coated RAL 9010 03 white...

Page 6: ...ss the switch rail down the table is lifted Lift the switch rail up the table is lowered Here too brief activation double tap must take place beforehand see section 7 Alternatively the switch rails ca...

Page 7: ...table feet all four castors are activated simultaneously The following moving options exist Stage 1 The castors of the table are locked and can neither be moved nor rotated Stage 2 centre position The...

Page 8: ...de guard with the necessary caution Never operate the side guard if the hands fingers etc of another person are located between the bars or on the mechanism of the side guard Risk of being crushed pin...

Page 9: ...dle part length mm 480 480 Foot part length mm 920 920 Weight approx depending on equipment kg 90 90 Min max height mm 500 to 940 490 to 940 Adjustment time motor s 22 Head part adjustment range 40 30...

Page 10: ...ion 3 7 A Protection class II protective insulation connecting cable without protective conductor Protection rating IPX4 resistant to water splashes Degree of protection B The motor is maintenance fre...

Page 11: ...nt block the foot levers The power plug must not come into contact with water or cleaning agents The electrical components must not display any external damage through which liquid could enter Do not...

Page 12: ...use Both can be replaced without too much effort Please request the corresponding installation diagram if required Replacement parts can be obtained directly from Dewert 7 Safety devices Tables with...

Page 13: ...ckaging Safety notice Pay attention to sharp edges and pointed objects during disposal The packaging materials produced are mainly o Cardboard Paper o Plastic o Wood when delivered on a pallet Please...

Page 14: ...ith the options of movability not wheel lifting fixing mechanism side guards and push handle the intended purpose of the table is extended and also provides for the ideal positioning of patients for t...

Page 15: ...damaged All switches and supply lines undamaged Function check With electromotive adjustment Move all motors with the foot switch or the hand switch to both limit positions until automatic switch off...

Page 16: ...ssories Accessories such as belts belt pads belt guides paper roll holders armrests etc undamaged and is proper and secure fixing function possible All necessary toggle screws present Possible movabil...

Reviews: