Dewert User Manual DX1 Series– Updated 05.2021/1 MDR                                                                     Page 10 of 16 

The motor is equipped with a thermal protective switch as standard. This causes the motor to cut out if the weight 
load is too extreme 

or the motor’s duty cycle (DC: 25 s/400 s) is exceeded. The table should be ready for use again 

after a 15-minute rest period. There is, therefore, no risk of overloading the motor. 

 

 
The mechanics of the tables have been designed with extensive safety margins in mind.  
The max. patient weight is provided with a 4-fold static safety factor, i.e. the design has been fully tested for a 4-
fold load. 
 

 

4.2 Technical data for electric motor 

 

Manufacturer:   

 

Hanning Elektro-Werke GmbH & Co, D-33813 Oerlinghausen 

Type of motor:   

 

SL 95  

Type of actuator: 

 

Brushless asynchronous industrial motor 

Mode of operation: 

 

Electromechanical linear motor with maintenance-free lifetime lubrication 

 

 

 

 

Intermittent duty 

– installed thermal switch 

 

 

 

 

Electronic activation with internal supply for the control element 

 

 

 

 

Duty cycle (DC): 25 s / 400 s 

 

 

 

 

i.e. move the table for max. 25 seconds under nominal load, then observe  

 

 

 

 

a break of at least 400 seconds.  

 

 

Nominal voltage: 

 

220 

– 240 V, 1-50/60 Hz 

Nominal consumption:    

850 W 

Current consumption:   

3.7 A 

Protection class: 

 

II (protective insulation), connecting cable without protective conductor 

Protection rating: 

 

IPX4 

– resistant to water splashes,  

Degree of protection: 

 

 
The motor is maintenance free. 
 
When operated with sinusoidal alternating voltage, the motors used do not cause field or line-borne interference 
within the meaning of EN 50081, Part 1 and 2, nor can their function be impaired by electromagnetic influences 
within the meaning of EN 50082, Part 1 and 2. 

 
 

4.3 Technical data for hydraulic system 

 

Manufacturer:   

Power-Packer Europa B.V., NL-7575 AT Oldenzaal 

Type:   

 

Compact MK5 long 

Mode of operation: 

Hydraulic cylinder with pump 

 
The hydraulic unit is maintenance free. 

 

 

5. Cleaning instructions 

 

Upholstery covering 

 
There are two different collections of upholstery covering to choose from: 
 

 

Skai Pandoria Plus (manufactured by Hornschuch/Continental in Germany) 

 

Skai Toronto EN (manufactured by Hornschuch/Continental in Germany) 

 

 

For cleaning and disinfection, a selection of agents from various manufacturers has been tested for 
compatibility. Please refer to the separate enclosed sheet. 
The table is equipped as standard with the 

Dewert hygiene standards

, which enable optimal cleaning and 

disinfection:

 

 

 

Hinge lids are made of the identical upholstery covering 

 

All undersides of the upholstery parts are coated with an upholstery covering and can thus be cleaned 
and disinfected 

Summary of Contents for 1000/H

Page 1: ...pliers guarantees years of trouble free operation provided the device is handled correctly and as intended in accordance with the conditions described in the user manual In the unlikely event you need...

Page 2: ...al number 5 2 Operating manual 6 2 1 Table design 6 2 2 Height adjustment 6 2 3 Upholstery adjustment 6 2 4 Movability model dependent or optional 7 2 5 Additional equipment 7 3 Additional accessories...

Page 3: ...ing the device We recommend that you keep these instructions for use near the device for future reference The device may only be used by authorised instructed and competent persons who are sufficientl...

Page 4: ...level flat and solid surface Before use activate the brakes on the castors and make sure they are working properly 1 6 Commissioning The device is ready for use upon delivery Remove the power cable fr...

Page 5: ...nly take hold of the underframe NOT the upper frame 1 8 Model designation and type labelling The exact model designation depends on the choice of frame colour 00 white powder coated RAL 9010 03 white...

Page 6: ...ss the switch rail down the table is lifted Lift the switch rail up the table is lowered Here too brief activation double tap must take place beforehand see section 7 Alternatively the switch rails ca...

Page 7: ...table feet all four castors are activated simultaneously The following moving options exist Stage 1 The castors of the table are locked and can neither be moved nor rotated Stage 2 centre position The...

Page 8: ...de guard with the necessary caution Never operate the side guard if the hands fingers etc of another person are located between the bars or on the mechanism of the side guard Risk of being crushed pin...

Page 9: ...dle part length mm 480 480 Foot part length mm 920 920 Weight approx depending on equipment kg 90 90 Min max height mm 500 to 940 490 to 940 Adjustment time motor s 22 Head part adjustment range 40 30...

Page 10: ...ion 3 7 A Protection class II protective insulation connecting cable without protective conductor Protection rating IPX4 resistant to water splashes Degree of protection B The motor is maintenance fre...

Page 11: ...nt block the foot levers The power plug must not come into contact with water or cleaning agents The electrical components must not display any external damage through which liquid could enter Do not...

Page 12: ...use Both can be replaced without too much effort Please request the corresponding installation diagram if required Replacement parts can be obtained directly from Dewert 7 Safety devices Tables with...

Page 13: ...ckaging Safety notice Pay attention to sharp edges and pointed objects during disposal The packaging materials produced are mainly o Cardboard Paper o Plastic o Wood when delivered on a pallet Please...

Page 14: ...ith the options of movability not wheel lifting fixing mechanism side guards and push handle the intended purpose of the table is extended and also provides for the ideal positioning of patients for t...

Page 15: ...damaged All switches and supply lines undamaged Function check With electromotive adjustment Move all motors with the foot switch or the hand switch to both limit positions until automatic switch off...

Page 16: ...ssories Accessories such as belts belt pads belt guides paper roll holders armrests etc undamaged and is proper and secure fixing function possible All necessary toggle screws present Possible movabil...