EN - 28
A-DV64
Filter Specifications
Standard Filter ............................................................................................................................................................................................................................................> 3.0 micron particles
Optional Fine Particle Filter ........................................................................................................................................................................................................................> 0.3 micron particles
Dynamic Short-Term Pressure Accuracy with and without DV6HH Humidifier (Per ISO 17510-1:2007)
Waveform
Volume [mL]
Breath Rate [min-1]
Set Pressure
[cmH
2
O]
Pressure Accuracy
CPAP only
[Pk-Pk cmH
2
O]
Pressure Accuracy
CPAP with Humidifier
[Pk-Pk cmH
2
O]
Sinusoidal cycle with
I:E = 1:1
500
10, 15, 20
3.0
0.5
0.5
7.0
0.5
0.5
12.0
1.0
1.0
16.0
1.0
1.0
20.0
1.0
1.0
Maximum Flow-Rate (per ISO 17510-1:2007)
Test Pressures
3.0 cmH
2
O
7.0 cmH
2
O 12.0 cmH
2
O 16.0 cmH
2
O 20.0 cmH
2
O
Measured pressure at the patient connection port (cmH
2
O)
2.9
6.6
11.4
15.0
19.0
Average flow at the patient connection port (l/min)
78.2
123.5
166.8
157.7
138.0
Long Term Static Pressure Accuracy .....................................................................................................................................................................................................................+/- 0.5 cmH
2
O
Additional Specifications
Equipment classification with respect to protection from electric shock ........................................................................................................................................................................... Class II
Degree of protection from electric shock .................................................................................................................................................................................................... Type BF Applied Part
Degree of protection against ingress of liquids ................................................................................................. IP21 Ingress Protection – Protected against finger access to hazardous parts;
protected against vertically falling water drops
Mode of operation ....................................................................................................................................................................................................................................................... Continuous
Equipment not suitable for use in the presence of a flammable anesthetic mixture with air or with oxygen or nitrous oxide. See instructions for use with supplemental oxygen.
US Patents
SmartCode Technology is protected under ...................................................................................................................................................................................................US Patent 8649510
Approvals
Meets RTCA/DO-160 .................................................................................................................................................... DO-160 Section 21 Category M for battery operation only Airline Use