EN - 17
A-DV64
PATIENT MESSAGES
Various messages may be displayed during your therapy; some include an audible alert
that is controlled by Master Volume Setting.
1. Notifications
These are friendly reminders to clean or replace components and, if enabled, are
generated as follows. Dismiss from Display by Button Press or Knob Movement.
Displayed Notification
Recurrence interval
Air Intake Filter Due for Cleaning
50 hours of use (while breathing)
Mask, Tubing, and Chamber Due for Cleaning
35 hours of use (while breathing)
2. Messages
These identify conditions that may require action by you/your provider or are
informational only. Dismiss from Display by Button Press or Knob Movement.
Displayed Message
Condition for Occurrence
Mask Off Please Check Mask Fit
10 seconds after mask off condition detected.
Auto-STOP Detected
When Auto-STOP occurs.
Auto-START Detected
When Auto-START occurs.
Wake-Up Alert <current time>
When Wake-Up Alert activates (alarm clock).
Error with Code, Prescription Not
Updated
When SmartCode Rx is entered but incorrect.
Card Detected Please Wait …
When SD card is detected. Do not remove SD
card while this message is displayed.
Card Removed Send to Provider
When SD card is removed after data transfer.
Card Error Contact Provider
If SD card error condition detected.
Settings Not Updated Contact
Provider
When prescription change via SD Card fails.
Card Transfer Successful
When valid SD Card inserted and data transfer
complete.
Card Detected Ready for use
When valid SD Card is inserted.
Firmware Update Please Wait …
When SD Card containing firmware update is
inserted.
3. Add-On Device Messages
These messages may be displayed when Add-On items are in use, Dismiss from Display
by Button Press or Knob Movement.
Displayed Message
Condition for Occurrence
Oximeter Logging Started
SpO2 _____
Pulse Rate _____
When Oximeter first detected and logging
starts
Oximeter Logging Stopped
When Oximeter signal is lost and logging
stops
Oximeter Finger Probe Off
When Oximeter finger probe off condition
detected
Oximeter Finger Probe On
SpO2 _____
Pulse Rate _____
When Oximeter finger probe off condition cor-
rected
4. Service Codes
These are divided into Critical or Non-Critical. Critical service codes put the unit in a fail-
safe (blower off) state.
Displayed Message
Action
Critical Service Code E0X - Where (X) is a
number or letter
NOTE – Code E01 is always shown in
English since language setting is unknown
Contact your equipment provider to
return unit for service. Message remains
on Display and cannot be dismissed.
Service Code E8X (non-critical - Where (X)
is a number or letter
Contact your equipment provider.
Dismiss from Display by Button Press or
Knob Movement.
NOTE- Service Code alerts are unaffected by notification enable/disable setting.