A-650
S p e c i f i c a t i o n s
13
SPECIFICATIONS
Size
3650 Series................................................................7.7" W x 4.1" H x 13" D
(19.5 cm x 10 cm x 33 cm)
4650 Series ..................................................................7" W x 3.8" H x 13" D
(18 cm x 10 cm x 33 cm)
5650 Series ...........................................................10.1" W x 10.5" H x 6.5" D
(26 cm x 27 cm x 17 cm)
Weight
3650 Series.....................................................................5.4 lbs. (2.45 kg)
4650 Series.....................................................................5.3 lbs. (2.40 kg)
5650 Series ......................................................................7.1 lbs. (3.2 kg)
Electrical
Requirements
3650D & 3650T ...........................................2.3 amps, 115 VAC 60 Hz
3650I ........................................................1.4 amps, 230 VAC 50 Hz
3650P ........................................................1.4 amps, 230 VAC 60 Hz
4650D, 46502, 4650AP, 4650L, 4650LD & 4650C ....2.5 amps, 115 VAC 60Hz
4650I.............................................................1.4 amps, 230 VAC 50 Hz
5650D, 5650H & 5650C ..................................1.3 amps, 115VAC 60Hz
5650I ..........................................................0.8 amps, 230VAC 50Hz
5650P .....................................................0.8 amps, 230VAC 50/60Hz
Sound Level
3650 Series ...........................................................................58 dBA
4650 Series............................................................................53 dBA
5650 Series............................................................................51 dBA
Power Consumption
3650D & 3650T..........................................140 watts maximum
3650I & 3650P ..........................................160 watts maximum
4650D, 46502, 4650AP, 4650L, 4650LD & 4650C .....140 watts max.
4650I.......................................................160 watts maximum
5650D, 5650H & 5650C ................................90 watts maximum
5650I & 5650P ..........................................100 watts maximum
Max. Compressor
Pressure
3650 & 4650 series......................................................35 psig or greater*
5650 series.................................................................30 psig or greater*
Compressor Free Air Flow
(3650 & 4650)......................................8 lpm or greater*
(5650).................................................9 lpm or greater*
Nebulizer Operating Pressure
all models ................................................11-16 psig
Nebulizer Liter Flow
all models.................................................................5-7 lpm
Nebulizer Capacity
all models .................................................................6 ml (cc)
Nebulizer Nebulization Rate
all models .......................................0.15 ml/min or greater
MMAD
all models.....................................................................Less than 5 microns
*Conditions may vary based on altitude above sea level, barometric pressure changes.
C l e a n i n g
A-650
12
COMPRESSOR CLEANING
1. With power switch in the "Off" position, unplug power cord from wall outlet.
2. Wipe outside of the compressor cabinet with a clean damp cloth every few days
to keep dust-free.
CAUTION–
Do not submerge in water; doing so will result in compressor damage.
FILTER CHANGE
1. Filter should be changed every 6 months or sooner if filter turns completely
gray in color.
2. Remove filter cap by grasping it firmly and pulling out of the unit
(Figs. A, B or C)
.
3. Remove dirty filter with fingers and discard.
4. Replace with a new DeVilbiss filter (part #5650D-602). Additional filters should
be purchased from your Sunrise provider.
5. Push filter cap with new filter into position.
CAUTION–
Reusing filter or substituting any other material such as cotton for a
DeVilbiss air-inlet filter will result in compressor damage.
NOTE–
An extra filter is packaged with the 4650 & 5650 series units.
MAINTENANCE
All maintenance must be performed by a qualified Sunrise provider or authorized ser-
vice center.
DANGER
Electric shock hazard. Do not remove compressor cabinet. All disassembly and
maintenance must be done by a qualified Sunrise provider.