ELECTRODE
GEOMETRY
LPS Electrode/ VAPR S
90
Electrode/Premiere90 Electrode
VAPR SYSTEM
VAPR;
VAPR II
VAPR
3
VAPR VUE
Default mode
BV2
V3
V3
Adv
Basic
Default
Vaporization
Power (W)
160
240
240
7
Default
Desiccation
Power (W)
120
120
120
5
Maximum
Vaporization
Power (W)
200
260
260
8
Maximum
Desiccation
Power (W)
160
160
160
6
Minimum
Cannula Size
5.5mm ID cannula
ELECTRODE
GEOMETRY
S
50
Electrode
VAPR SYSTEM
VAPR;
VAPR II
VAPR
3
VAPR VUE
Default mode
V2
V2
V2
Adv
Basic
Default
Vaporization
Power (W)
60
60
60
3
Default
Desiccation
Power (W)
45
45
45
2
Maximum
Vaporization
Power (W)
90
90
90
4
Maximum
Desiccation
Power (W)
60
60
60
3
Minimum
Cannula Size
Not Applicable, Stab incision only
ELECTRODE
GEOMETRY
Premiere50 Electrode
VAPR SYSTEM
VAPR;
VAPR II
VAPR
3
VAPR VUE
Default mode
BV2
V2
V2
Adv
Basic
Default
Vaporization
Power (W)
60
160
160
6
Default
Desiccation
Power (W)
45
45
45
2
Maximum
Vaporization
Power (W)
90
200
200
7
Maximum
Desiccation
Power (W)
60
60
60
3
Minimum
Cannula Size
Not Applicable, Stab incision only
DURING SURGERY
1. Close suction outflow of the electrode prior to
insertion of the electrode.
2. Insert the electrode into a fluid-filled joint space through
a prepared entry point. Ensure that the electrode is
surrounded by conductive irrigant solution when in use.
3. Adjust the flow rate, by opening suction outflow, to
optimize the fluid removal rate and bubble extraction.
4. Reduce suction when device is not in use.