8 Maintenance
8.3 Color calibration
Dentsply Sirona
Operating Instructions (valid for Canada)
72
66 37 487 D3652
D3652.201.01.21.02 03.2021
Carry out a color calibration also after changing a mirror sleeve.
Heavily scratched mirror sleeves may not be used for a color analysis.
Storing a color-calibration set
The color-calibration set must be stored in its packaging in a dry place
which is protected from light. It must be used with a disinfected
Omnicam as the color-calibration set must itself not be disinfected. If
dust accumulates on the inside of the color-calibration set, it must be
carefully removed using compressed air.
Switch on the color analysis
1.
In the software, navigate to the system menu and click on the
"Configuration" button.
2.
Click on the
"Devices" button.
3.
Click on the
"Omnicam" button.
4.
Select the
"Shade Detection" option.
- You can choose between various color systems (
"Shade Guide
Selection").
- You can decide whether you would like to be notified in 14 days
when the color calibration is needed again.
- If color analysis is not possible with your camera, a corresponding
notice will appear. A color calibration is also not available in this
case.
5.
Confirm the changes below with
"Ok".
6.
Click the
"Color Calibration" button and carry out the color
calibration.