6.5 Commissioning
66 37 487 D3652
D3652.201.01.21.02 03.2021
37
English
Dimensions of the PC as of hardware version 5.4.1
Width
Height
Depth
in mm
230
360
380
in inches
9
1
/
16
14
11
/
64
14
61
/
64
A
PC
D
Ventilation slots in the cabinet
B
Cabinet
E
PC power supply unit/PC air
extraction
C
Cabinet fan
F
PC air intake
NOTE
The temperature inside the cabinet affects the performance of the PC.
Full performance can only be guaranteed up to a temperature of 35°C
(95°F) (measured at the air intake of the PC). A drop in performance
can become apparent if the scanning process slows down when the
temperature rises above 35°C (95°F).
- The cabinet should have an active ventilation or air extraction
system; see the sketch for an example.
- It is particularly important to check the temperature inside the
cabinet when the ambient temperature is higher (in summer, for
example). If necessary, open the door or place the PC somewhere
else.