Section VII User Service Information
Arm System
Symptom
Possible Cause(s)
Solution
Horizontal arms
drift left or right
Unit is not properly leveled
Verify unit is level. Also, verify arms are properly seated. Adjust
tension or set screws as needed.
Unit head is not
level from front
to back.
Pivot is not properly adjusted
Remove end cap cover and adjust bolt as necessary.
Adjustable arm
does not hold
position.
Improper adjustment
With nothing on delivery head, adjust bolt at the rear of the
fl
ex arm so
unit stays in place.
Faulty brake switch
Make sure air is running through the brake toggle when the brake is on.
No air to brake switch
Ensure unit master toggle is in ON position.
Delivery Head
Delivery Head
Cuspidor (Optional)
Symptom
Possible Cause(s)
Solution
Bowl rinse has no
water
Stop valve closed
Open stop valve inside USC compartment.
Pinched or restricted 1/4" green tubing from
fl
oor utility box to post-mounted console
Check for restricted or obstructions and free accordingly.
Water actuator valve in
fl
oor utility box is not
activated
Verify unit master switch is turned ON.
Verify air signal is present at valve on water line.
Verify regulator is open and pressure is at 40 PSI.
Bowl rinse button not working
Verify air signal is present at the bowl rinse poppet valve on the
cuspidor when the bowl rinse button is depressed.
Bowl rinse actuator valve in cuspidor is not
opening
Verify air signal is present at actuator valve when button is depressed.
At actuator valve in cuspidor, check for water supply from
fl
oor utility
box. Also, ensure water passes through the valve when the rinse
button is depressed.
Bowl rinse runs
continuously
Water pressure is too high
Set water pressure gauge at 40 PSI.
Bowl rinse actuator valve is stuck open
Check the valve in the cuspidor for air signal. If air is present, check
bowl rinse poppet valve. If air is not present, replace actuator valve.
Bowl does not
drain or drains
slowly
Restricted or pinched drain line
Examine drain line and free and obstructions
Restricted or pinched vent line
Examine vent line and free any obstructions.
No water
fl
ows
from cup to
fi
ll
Cup
fi
ll button is not working
Check actuator valve in cuspidor for air signal at cup
fi
ll actuator valve
in post-mounted console when button is depressed.
Check for pinched airlines.
Cup
fi
ll actuator valve in cuspidor is not open Verify signal is present at actuator when cup
fi
ll button is depressed. If
air and water are present, replace valve.
Troubleshooting
(Continued)
Summary of Contents for Core
Page 1: ...CORE TM Delivery Unit Magellan Mount User Manual ...
Page 3: ...www DentalEZ com 866 DTE INFO ii CORETM Magellan Unit PN 2717 267C ...
Page 11: ...www DentalEZ com 866 DTE INFO 8 CORETM Magellan Unit PN 2717 267C ...
Page 15: ...www DentalEZ com 866 DTE INFO 12 CORETM Magellan Unit PN 2717 267C ...
Page 31: ...www DentalEZ com 866 DTE INFO 28 CORETM Magellan Unit PN 2717 267C ...
Page 39: ...www DentalEZ com 866 DTE INFO 36 CORETM Magellan Unit PN 2717 267C ...