11. Place a level on the Magellan unit support,
parallel with and perpendicular to the chair
center line, then check the level of the unit
support.
NOTE:
If leveling is necessary, adjust the 3/8" set
screws until the unit support is level
(Figure 12)
.
12. Tighten all hardware (55 foot-pounds for 1/2"
bolt).
Figure 12. Adjust 3/8" set screws if leveling is necessary
1/2" Bolts
Leveling Set Screws
Magellan Unit Support
(Continued)
Section III Installation
Umbilical Assembly
CORE Chair
NOTE:
Provide su
ff
icient slack in the umbilical
assembly to avoid stretching when the chair is in
its full
UP
position.
1. From the umbilical bag of supplies, select
the umbilical assembly with the supplied
L-bracket.
2. Route the umbilical between the hydraulic
pump and the control board, then under the
cantilever arm.
3. Route the remainder of the umbilical through
the pivot and up into the chair casting, using
tie wraps to secure.
NuSimplicity Chair
1. From the umbilical bag of supplies, select
the umbilical assembly with the supplied
L-bracket.
2. Route the umbilical between the hydraulic
pump and the control board, then under the
cantilever arm.
3. Tie wrap the umbilical to the bracket
(Figure 13)
.
For tubing connections, see the tubing
diagram in the back of this user manual.
NOTICE
Figure 13. Tie wrap umbilical to L-bracket
Tie Wrap
L-bracket
Summary of Contents for Core
Page 1: ...CORE TM Delivery Unit Magellan Mount User Manual ...
Page 3: ...www DentalEZ com 866 DTE INFO ii CORETM Magellan Unit PN 2717 267C ...
Page 11: ...www DentalEZ com 866 DTE INFO 8 CORETM Magellan Unit PN 2717 267C ...
Page 15: ...www DentalEZ com 866 DTE INFO 12 CORETM Magellan Unit PN 2717 267C ...
Page 31: ...www DentalEZ com 866 DTE INFO 28 CORETM Magellan Unit PN 2717 267C ...
Page 39: ...www DentalEZ com 866 DTE INFO 36 CORETM Magellan Unit PN 2717 267C ...