MAINTENANCE - 7
7.4
Replacing the bacteriological filter
NOTE
Replace the bacteriological filter with the device switched off.
Unscrew the bacteriological filter and replace it with a new one. Install the filter in its housing, taking care to ensure that it
is fully screwed in.
7.5
Cleaning the instruments before sterilization
To ensure the maximum duration and reliability of the device, it is advisable to take the utmost care over the cleaning and
washing of instruments.
One of the main causes of premature wear of the sterilizer is due to residues from imperfectly cleaned instruments, resulting
in the formation of stains, limescale and progressive clogging of filters, solenoid valves and hydraulic circuits.
7.6
Schedulled maintenance
Pos.
Check/Activity
Maintenance
Special maintenance *
1
Door adjustment
1 Year / 1.500 cycles
- Gasket replacement
1 Year / 1.500 cycles
- Disk-door clearance check
1 Year / 1.500 cycles
- Closing force check
1 Year / 1.500 cycles
- Lubrication
1 Year / 1.500 cycles
- Component wear check
1 Year / 1.500 cycles
- Closing pin replacement
4 Years / 10.000 cycles
- Screw hinge replacement
4 Years / 10.000 cycles
- Screw tightening
4 Years / 10.000 cycles
2
Calibration / Validation
1 Years
- Altitude setting check
1 Years / 1.500 cycles
3
Filter cleaning / replacement
1 Years / 1.500 cycles
- Bacterial filter replacement
6 Months / 500 cycles
4
Pump feature check
1 Years / 1.500 cycles
- Pump replacement
10.000 cycles
5
Tank cleaning
1 Years / 1.500 cycles
6
Condenser cleaning
1 Years / 1.500 cycles
7
Safety valve replacement
4 Years / 10.000 cycles
*) Extraordinary maintenance must be performed by qualified personnel authorized by Dental X. If the sterilizer needs to be sent back or
removed to carry out repairs in a workshop or factory, remember to enclose a photocopy of the sales invoice together with the sterilizer
and, in the case of a return, the return merchandise authorization (RMA), which must always be requested in advance from the Dental X
customer care office before sending the device.
The purchased sterilizer meets the relevant requirements of current safety standards. The parameters set by the manufacturer are designed
to ensure the sterility of the load under the conditions specified in the manual.
Read this manual carefully before using the sterilizer; improper use may terminate or invalidate the warranty conditions applied to the
purchase of the device.
NOTE: For any questions, requests or inquiries regarding the device, first contact the dealer who supplied the product.