INSTALLATION - 4
4.3
Initial start-up
WARNING
The following operations must be carried out by qualified and properly trained personnel.
Incorrect procedures and settings can jeopardize the quality of sterilization and cause hazards.
Check that the power supply has the right voltage and plug the power cord
into the outlet.
Fill the demineralized water tank up to the maximum level. The tank holds
approximately 4 litres of water. The minimum water level LED switches off,
indicating that the tank is filling. When the full tank indicator lights up, this
indicates that the tank has been properly filled.
CAUTION
The use of poor quality demineralized water can leave calcium deposits on the instruments, on the
inside of the chamber and on the trays. Read the label on the distilled water container carefully. Do
not use domestic tap water, even if treated with a filter or water softener.
WARNING
Do not use battery water or other fluids or additives, as these can cause irreversible damage to
the device and hazards to the operator.
Switch the device on by the main switch. For daily inactive periods, the main
switch can be left in the ON position, as the power consumption in stand-by
is almost nil.
Remove the rack and trays from the chamber and close the door.
NOTE
The door remains locked when the device is switched off; if it is still locked when switched on, turn
the device off and then on again.