![dental X dxp DOMINA PLUS B Operator'S Manual Download Page 30](http://html.mh-extra.com/html/dental-x/dxp-domina-plus-b/dxp-domina-plus-b_operators-manual_2483369030.webp)
30
USING THE STERILIZER - 5
5.6
Program selection
Various types of programs are available to the operator, depending on the degree of sterilization desired, with a class B or
class S sterilization cycle, as described in the program table.
To select programs
B134, B121 and S134
, press the corresponding button on the operator panel.
To select one of the
CUSTOM
programs, press the CUSTOM button on the operator panel, then the
or
buttons to
choose from the available
S1, S2, S3, S4 and S5
options.
The
S1, S2, S3 and S4
programs are preset by the manufacturer;
S5
can be customized by the operator (see the paragraph
on Setting the S5 Custom program).
When the desired program is selected, the displays show the program settings for 5 seconds.
PROGRAM TABLE
Program
Parameter
Load
Cycle
Process
values
Maximum
Load
B134
134°C - 5‘
3 pre-vacuum phases
drying 10’ (6' with 4’ with
ventilation)
Solid, porous,
hollow type A
and B wrapped
Class B
134-137°C
2.04-2.25 bar
4 kg solid or
1.5kg porous
B121
121°C - 20’
3 pre-vacuum phases
drying 11’ (7' with 4’ with
ventilation)
Class B
121-124°C
1.04-1.24 bar
S134
134°C - 3’
2 pre-vacuum phases
drying 5’ (3' with 2’ with
ventilation)
Unbound solid
instruments
Class S
134-137°C
2.04-2.25 bar
4 kg solid
Custom
S1
105°C - 8’
(3 pre-vacuum phases, drying 6’+ 4’)
Solid, porous,
hollow type A
and B
Disinfection
105-108°C
0.21-0.35 bar
4 kg solid or
1.5kg porous
Custom
S2
134°C - 5’
(4 pre-vacuum phases, drying 7’+ 5’)
Solid, porous,
hollow type A
and B wrapped
Class B
134-137°C
2.04-2.25 bar
Custom
S3
121°C - 20’
(4 pre-vacuum phases, drying 7’+ 5’)
Class B
121-124°C
1.04-1.24 bar
Custom
S4
134°C - 18’
(3 pre-vacuum phases, drying 6’+ 4’)
Prion
Class B
134-137°C
2.04-2.25 bar
Custom
S5
Settable parameters
Temp: 105 - 135°C
Time: 3’- 90’
Pre-vacuum phases: 2, 3 or 4
Drying with ventilation:
3’+2’, 6’+4’, 8’+6’
Loading
according
to the set
parameters
According
to the set
parameters
105-138°C
0.21-2.30 bar
Helix Test
Bowie&Dick
Test
134°C - 3’ 30’’
3 pre-vacuum phases
drying 10’ (6' with 4’ with
ventilation)
Helix Test
(HTS100 class 5
cod. 9900051)
B&D Test
(3M
TM
COMPLY
TM
cod. 1300)
TEST
134-137°C
2.04-2.25 bar
Helix test pack
or equivalent
B&D test pack
or equivalent
Vacuum
Test
Temperature below 35°C
TEST
< 35°C
Empty
chamber