36
USING THE STERILIZER - 5
CAUTION
A negative result from this check does not prevent use of the sterilizer; however, it is recommended
to replace demineralized water of less-than-ideal quality with better quality water.
The following table shows the recommended minimum parameters for the water to be used
Pollutants
Supply water
Condensate
evaporation residue
≤ 10 mg/l
≤ 1.0 mg/l
silicon oxide
≤ 1 mg/l
≤ 0.1 mg/l
iron
≤ 0.2 mg/l
≤ 0.1 mg/l
cadmium
≤ 0.005 mg/l
≤ 0.005 mg/l
lead
≤ 0.05 mg/l
≤ 0.05 mg/l
residues of heavy metals, excluding iron,
cadmium and lead
≤ 0.1 mg/l
≤ 0.1 mg/l
chlorine
≤ 2 mg/l
≤ 0.1 mg/l
phosphates
≤ 0.5 mg/l
≤ 0.1 mg/l
Conductivity (at 20°C)
≤ 15 uS
≤ 3 uS
Ph
from 5 to 7.5
from 5 to 7
appearance
Colourless, clean without
sediment
Colourless, clean without
sediment
hardness
≤ 0.02 mmol/l
≤ 0.02 mmol/l
5.12 Connections
5.12.1 Connection to an external printer
The device does not have an integrated printer, but is designed for connection to an external printer, to which the process
data is sent in order to document and certify the sterilization.
The use of a printer, which is required in some countries, is destined to become ever more frequent, in order to also ensure
the proper sterilization of dental materials from a medical-legal perspective.
The printer cable should not be more than 3 metres in length.
A printer with RS232 serial interface can be connected to the serial interface port on
the device.
In sequence:
1. turn on the printer;
2. turn on the sterilizer.
The report is printed automatically during the cycle and will contain the following
information:
• date and time of the process
• serial number of the cycle
• cycle and parameters selected
• type of process: sterilization or disinfection
• start and end times of the sterilization phase
• time of completion of the drying phase.
In the case of malfunction or interruption of the cycle, the printer will provide the message
CYCLE ABORTED - NOT STERILE
and
will indicate the type of alarm encountered.
At the end of the working day, turn the printer off.
To program the print report language, see the paragraph on “Setting temperature and pressure measurement units and
selecting the language”.
The PRINTER port only interfaces directly with the printer.
The same above described procedure can be used for the connection to the CUSTOM
printer
The CUSTOM printer is enabled to the normal print of the report on thermal paper and/or
to the barcode printing using the adequate label roll.