![dental X dxp DOMINA PLUS B Operator'S Manual Download Page 21](http://html.mh-extra.com/html/dental-x/dxp-domina-plus-b/dxp-domina-plus-b_operators-manual_2483369021.webp)
21
INSTALLATION - 4
4.2
Positioning
Check that the power supply voltage to the device matches that shown on the
regulatory label on the rear panel, that the power outlet is designed to supply at
least 16A and that it has an earth connection.
If the installation makes the main power switch inaccessible, a specially dedicated
accessible electrical disconnection switch should be provided.
WARNING
The manufacturer is not liable for damage to property or persons caused by electrical
systems that are unsuitable or have no earth connections.
The device must be installed on a flat surface. If the support surface
is perfectly horizontal, the front feet are already adjusted with a slight
inclination to facilitate the flow of water during draining.
If the support surface is not perfectly level, adjust the front legs, raising
or lowering them to obtain a slight inclination, as shown in the figure.
WARNING
Do not place the device on a fragile surface that could be damaged or cause fire or smoke if
hot objects fall.
For proper operation, leave a free space of at least 4 cm between the rear
of the device and any wall.
Do not install the device near heat sources or in damp or poorly ventilated
spaces. The room must provide air circulation with at least 10 air changes
per hour; a recirculating air ventilation system (e.g., an electric fan) cannot
be used as an alternative.
CAUTION
The safety valve is on the rear of the device. When triggered by excess pressure, it releases very
hot steam into the environment. Position the device to avoid risk of burns to the operator.