28
Reservoir Volume
is zero
NEXT to continue
Reservoir Volume
Low
NEXT to continue
Error Detected
E
(
error code
)
Upstream Occlusion
Press STOP to stop
Press NEXT to restart
(Model 6101 only)
A l a r m s a n d M e s s a g e s
When you see:
Take this action:
Fluid is not flowing from the
fluid container to the pump.
Check for a kink, a closed
clamp, or an air bubble in the
tubing between fluid container
and pump. Press
⁄
to stop
the pump and silence the
alarm, then remove the ob-
struction and press
„
to
restart the pump.
The Reservoir Volume has
reached 0.0 ml. Press
„
to
stop the alarm. Then install a
new fluid container, if appro-
priate.
The Reservoir Volume value is
low, indicating that the level of
fluid in the fluid container is
low. Prepare to install a new
fluid container.
There is a problem with the
pump. Close the tubing clamp
and remove the pump from
service. Contact your clinician.