
16
2000-03-08 D. Z
«Prizm Hinge C
A t t a c h i n g a Cassette
WARNING:
• You must use a CADD
®
Extension Set with Anti-Siphon
Valve or a CADD
®
Administration Set with either an
integral or add on Anti-Siphon Valve to protect against
unregulated gravity infusion that can result from an
improperly attached cassette. Unregulated gravity infu-
sion can result in death or serious injury.
• For detailed instructions and warnings pertaining to the
Medication Cassette
™
Reservoir or CADD
®
Administration
Set, please refer to the instructions for use supplied with
those products.
• Frozen medication must be thawed at room temperature
only. Do not heat the Medication Cassette
™
Reservoir in a
microwave oven as this may damage the medication, the
Medication Cassette
™
Reservoir, or cause leakage.
Use aseptic technique as
instructed by your clinician.
1. Clamp the tubing on the
new Medication Cas-
sette
™
Reservoir or
CADD
®
Administration
Set.
2. Insert the cassette hooks
into the hinge pins on
the pump.
3. Place the pump upright on a firm, flat surface. Press down
so the cassette fits tightly against the pump.
®
Summary of Contents for CADD-Prizm VIP PCA 6100
Page 31: ...29...
Page 33: ...31 Notes...