12
Starting the Pump
When you start the pump, it will review the program, and
then begin delivering medication. If the pump will not start,
a message should appear on the display.
Refer to the Mes-
sages and Alarms section on page 25.
WARNING: Prior to starting infusion, inspect the fluid path
for kinks, a closed clamp, or other upstream obstructions,
and remove all air bubbles to prevent air embolism. An
undetected upstream occlusion may result in under- or non-
delivery of drug and, depending on the drug being adminis-
tered, could result in death or serious injury. Air embolism
can result in death or serious injury.
1. Press
⁄
.
2. When
Start the pump?
ap-
pears, press
´
.
Starting pump…
will appear. The
pump will automatically review
and display the preprogrammed
settings.
RUNNING
will appear on the
main screen, the green light will
blink, and fluid delivery will
begin as programmed by your
clinician.
Start the Pump?
Press Y or N
Starting pump...
*** PCA 6210X ***
RUNNING
Res Vol
50.0 ml
Press NEXT to advance