17
4. Insert the side of the key
or a coin into the latching
button, push in, and turn
counterclockwise until the
mark on the latch lines up
with the solid dot.
A message will appear on
the display showing the type
of cassette you have latched.
5. Insert the key into the lock and turn
counterclockwise until the white
mark lines up with the solid dot.
NOTE:
The cassette must be latched
and locked in order to start the pump.
WARNING: Attach the cassette (the part of the Medication
Cassette
™
Reservoir or CADD
®
Administration Set that
attaches to the pump) properly. An improperly attached or
detached cassette could result in unregulated gravity infu-
sion of medication from the fluid container or a reflux of
blood, which could result in death or serious injury.
6. Gently twist and pull on the cas-
sette to make sure it is firmly
attached.
®
®
®