25
9 volt Battery
Low
NEXT to continue
9 volt Battery
Depleted
Install good battery
Cassette Damaged
Free flow may occur
Clamp Tubing
NEXT to silence
(blank display or
random characters)
Alarms and Messages
If there are alarms or special messages you need to be aware
of, the pump will beep or sound an alarm. Look at the screen
and follow the steps in this table, or press
Å
for help.
When you see:
Take this action:
The 9 volt battery is low but the
pump is operable. Change the 9
volt battery soon.
Install a new 9 volt battery. The
pump will not run with a
depleted 9 volt battery. A good
battery must be installed even
when an external source of
power is connected.
A two-tone alarm is sounding
and the amber light stays on.
The 9 volt battery is depleted.
Install a new battery.
The pump detects the cassette is
damaged. Close the tubing
clamp and inspect the cassette
for damage. Replace it if
necessary.
Summary of Contents for CADD-Prizm VIP PCA 6100
Page 31: ...29...